British Journal of Ophthalmology 2003;87:1481-1486; doi:10.1136/bjo.87.12.1481
Copyright © 2003 by the BMJ Publishing Group Ltd.
British Journal of Ophthalmology 2003;87:1481-1486
© 2003 BMJ Publishing Group Ltd

EXTENDED REPORT

Is manual counting of corneal endothelial cell density in eye banks still acceptable? The French experience

G Thuret1, C Manissolle1, S Acquart2, J-C Le Petit2, J Maugery1, L Campos-Guyotat1, M J Doughty3 and P Gain1

1 "Cell Survival and Adhesion in Cancers and Graffs" Research Group, EA 3063, Faculty of Medicine, and Ophthalmology Department, Bellevue University Hospital, Saint-Etienne, France
2 Cornea Bank, Etablissement Français du Sang Auvergne-Loire, Saint-Etienne, France
3 Department of Vision Sciences, Glasgow-Caledonian University, Glasgow, UK

Correspondence to:
Correspondence to:
Gilles Thuret
Service d’ophtalmologie, Pavillon 50A, Hôpital Bellevue, CHU Saint-Etienne, F-42055 Saint-Etienne Cedex 2, France; gilles.thuret{at}univ-st-etienne.fr

Aim: To examine the differences in manual endothelial cell counting methods in French eye banks and to analyse whether these differences could explain some substantial discrepancies observed in endothelial cell density (ECD) for corneas made available for transplant.

Methods: A questionnaire was sent to the 22 eye banks asking for details of the technical features of the light microscopes used, the microscope calibration, strategy for cell counting, the technical staff, and the method of presenting endothelial data.

Results: All eye banks responded and 91% (20/22) used only manual counting methods, in real time, directly through a microscope, and 62 different technicians, with varying experience, were involved in such counting. Counting of cells within the borders of a grid that were in contact with two adjacent borders was the most common method (17/22, 77%). Of the eight banks (8/22, 36%) that did not calibrate their microscopes, six reported the highest ECD values. Of the 14 others (64%), six applied a "magnification correcting factor" to the initial cell counts. In five of these cases, the corrected ECD was lower than estimated on initial count. Most of the banks (12/22, 55%) counted 100 cells or less in one to six non-adjacent zones of the mosaic. 14 of the banks (14/22, 64%) also graded cell polymegethism while seven (7/22, 32%) also graded pleomorphism ("hexagonality").

Conclusions: Lack of microscope calibration appears to be the leading cause of variance in ECD estimates in French eye banks. Other factors such as differences in counting strategy, the evaluation of smaller numbers of cells, and the different extent of experience of the technicians may also contribute to intraobserver and interobserver variability. Further comparative studies, including cross checking and the outcome of repeated counts from manual methods, are clearly needed with cross calibration to a computer based image archiving and analysis system.

Keywords: cornea; organ culture; endothelium; quality control


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

This article has been cited by other articles:

  • Deb-Joardar, N., Thuret, G., Gavet, Y., Acquart, S., Garraud, O., Egelhoffer, H., Gain, P. (2007). Reproducibility of Endothelial Assessment during Corneal Organ Culture: Comparison of a Computer-Assisted Analyzer with Manual Methods. IOVS 48: 2062-2067 [Abstract] [Full Text]  
  • Deb-Joardar, N., Thuret, G., Racine, G.-A., Pons, D., Brun, G., Parriaux, O., Peoc'h, M., Acquart, S., Gain, P. (2006). Standard microlithographic mosaics to assess endothelial cell counting methods by light microscopy in eye banks using organ culture.. IOVS 47: 4373-4377 [Abstract] [Full Text]  
  • Thuret, G., Manissolle, C., Campos-Guyotat, L., Guyotat, D., Gain, P. (2005). Animal Compound-Free Medium and Poloxamer for Human Corneal Organ Culture and Deswelling. IOVS 46: 816-822 [Abstract] [Full Text]  

This Article

Services
Citing Articles
Google Scholar
PubMed
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

Ophthalmology Jobs

Ophthalmology Jobs