EXTENDED REPORT
Potential of the 1 CU accommodative intraocular lens
Department of Ophthalmology and Eye Hospital, Faculty of Clinical Medicine Mannheim, University of Heidelberg, 68167 Mannheim, Germany
Correspondence to:
Correspondence to:
Dr G Sauder
Universitäts-Augenklinik, Theodor-Kutzer-Ufer 1-3, 68167 Mannheim, Germany; gangolf.sauder{at}augen.ma.uni-heidelberg.de
Aim: To assess the accommodative power of a new foldable monofocal intraocular lens.
Method: A prospective randomised non-masked clinical interventional study. The study included 40 patients attending the hospital for cataract surgery and who were randomly distributed into a study group receiving a new foldable monofocal intraocular lens with flexible haptics, and a control group receiving a standard foldable intraocular lens. Mean follow up period was 8.51 (SD 1.34) months (range 411 months) Standard cataract surgery consisted of clear cornea incision, capsulorrhexis, phacoemulsification, and intraocular lens implantation, with topical anaesthesia. The main outcome measures were preoperative and postoperative visual acuity for near and distance; range of accommodation; change in anterior chamber depth.
Results: In the study group compared with the control group, range of accommodation was significantly (p = 0.01) higher (1.01 (SD 0.4) dioptres versus 0.50 (0.11) dioptres) and change in anterior chamber depth was significantly more pronounced (0.82 (0.30) versus 0.40 (0.32), p = 0.01). Both groups did not vary significantly in best corrected vision (0.94 (0.12) versus 0.93 (0.18); p = 0.74).
Conclusion: During a mean follow up period of 8 months after implantation, the new foldable monofocal intraocular lens with flexible haptics showed an accommodative power of about 1 dioptre, which was significantly higher than the accommodative power of a conventional monofocal flexible intraocular lens. The difference in the accommodative power between the two intraocular lenses was paralleled by a difference in the change of the anterior chamber depth.
Keywords: accommodation; intraocular lens; refractive cataract surgery
Relevant Article
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
Br. J. Ophthalmol. 2005 89: 1231.
This article has been cited by other articles:
-
Duncan, G., Wang, L., Neilson, G. J., Wormstone, I. M.
(2007). Lens Cell Survival after Exposure to Stress in the Closed Capsular Bag. IOVS
48: 2701-2707
[Abstract] [Full Text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.