British Journal of Ophthalmology 2007;91:497-499
EXTENDED REPORT
Accuracy and performance of a commercially available Dosing Aid
UPMC Eye Center, University of Pittsburgh, Pittsburgh, Pennsylvania, USA
Correspondence to:
Correspondence to:
Dr Robert J Noecker
University of Pittsburgh, 203 Lothrop Street, Suite 800, Pittsburgh, PA 15213, USA; noeckerrj{at}upmc.edu
Aim: To evaluate the recently introduced Travatan Dosing Aid (TDA) for its accuracy in recording and dispensing eyedrops.
Methods: The number of eyedrops dispensed with each lever depression and agreement of total number of drops dispensed with that recorded by the device was evaluated in a controlled setting.
Results: The TDA correctly recorded a drop being dispensed 100% of the time with full TDA lever depression for <3 s. Under these conditions, agreement between numbers of drops dispensed and recorded was 99%. However, failure to fully depress the lever or prolonged lever depression for >4 s resulted in unreliable TDA recording.
Conclusion: Eyedrops were reliably recorded by the TDA after each full lever depression. However, patients need to be instructed about optimal technique so that evaluation of compliance is not confounded by mechanical factors.
Abbreviations: TDA, Travatan Dosing Aid
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.