Impact of simulated light scatter on scanning laser Doppler flowmetry

Abstract

Aim: To determine the impact of simulated light scatter on scanning laser Doppler flowmetry (SLDF) assessment of retinal capillary blood flow and retinal image quality.

Methods: One eye of 10 normal subjects (mean (SD) age 24 (1.7) years, range 22–27) was randomly selected. Varying concentrations of polystyrene microspheres were suspended in optically clear cells to simulate light scatter. The microsphere concentrations used were 0.05%, 0.03%, 0.02%, 0.01%, and a cell containing only water. LogMAR visual acuity and contrast sensitivity were measured both with and without cells. Optimal focus and alignment was established by acquiring three SLDF images each of the optic nerve head (ONH) and of the macula using the Heidelberg retina flowmeter (HRF) with no cell in place. SLDF images were subsequently acquired with each of the light scatter cells mounted in front of the HRF. The group mean retinal capillary blood flow was compared using repeated measures analysis of variance (reANOVA) as a function of microsphere concentration.

Results: Retinal capillary blood flow increased significantly in the ONH, nasal macula, fovea, and temporal macula with increasing microsphere concentration (p<0.0001). Using Dunnett’s post hoc test, retinal capillary blood flow was found to be significantly increased relative to the no cell condition for the 0.03% and 0.05% cell concentrations.

Conclusions: Simulated light scatter produces an artifactual increase in retinal capillary blood flow. The impact of cataract on SLDF measurements has yet to be determined.

  • AFFPIA, automated full field perfusion image analysis
  • HRF, Heidelberg retina flowmeter
  • ONH, optic nerve head
  • SLDF, scanning laser Doppler flowmetry
  • light scatter
  • polystyrene microspheres
  • cataract
  • retina
  • capillary blood flow
  • AFFPIA, automated full field perfusion image analysis
  • HRF, Heidelberg retina flowmeter
  • ONH, optic nerve head
  • SLDF, scanning laser Doppler flowmetry
  • light scatter
  • polystyrene microspheres
  • cataract
  • retina
  • capillary blood flow

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.