rss
Br J Ophthalmol 1980;64:591-596 doi:10.1136/bjo.64.8.591
  • Research Article

Use of the Arden grating test for screening.

Abstract

One hundred and forty patients were referred for a clinical evaluation of their eyes, having produced high scores on screening by the Arden grating test (AGT). In these patients 66% of the eyes were found to be normal, 31% abnormal, and 3% indeterminate on clinical examination. Suitable criteria for referral for clinical examination were determined and the recommended scores are: (a) A total AGT score of greater than or equal to 80; (b) One or more plates scoring greater than or equal to 17; (c) A difference between the two eyes of greater than or equal to 13. With these criteria at least 95% of the abnormal eyes will be detected. As a screening test the AGT is equal to the Snellen distance visual acuity test at separating the normal group from the abnormal group, and on the recommended criteria the AGT gives a better ratio of false positives to false negatives.

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of BJO.
View free sample issue >>

Free archive
The full back archive is now available for BJO. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>