Br J Ophthalmol 1991;75:42-44 doi:10.1136/bjo.75.1.42
  • Research Article

Clinical evaluation of automated refraction in anterior chamber pseudophakia.

  1. P Sunder Raj,
  2. J R Villada,
  3. K Myint,
  4. A E Lewis and
  5. T Akingbehin
  1. Department of Ophthalmology, District General Hospital, Southport.

      Abstract

      Automated refraction with the Canon RK-1 Autoref keratometer was evaluated in 85 eyes of 85 patients with an anterior chamber intraocular lens implant and a best corrected visual acuity of at least 6/12. Autorefraction readings were obtained in 80 (94%) of these eyes. The agreement between autorefraction and clinical refraction data was 89% for spherical equivalence less than 0.51 dioptres (D), 91% for sphere power less than 0.51 D, 82% for cylinder power less than 0.51 D, and 91% for cylinder axis less than 11 degrees. Autorefraction provides acceptably accurate postoperative refraction values in anterior chamber pseudophakia.

      Register for free content

      The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.