Br J Ophthalmol 1992;76:303-306 doi:10.1136/bjo.76.5.303
  • Research Article

Calculation of the power of anterior chamber implants.

  1. J R Villada,
  2. P S Raj and
  3. T Akingbehin
  1. Department of Ophthalmology, District General Hospital, Southport.

      Abstract

      Ninety seven eyes with anterior chamber implants and a best corrected visual acuity of at least 6/12 were studied, firstly, to compare the predictive accuracies of the commonly used formulas for intraocular lens (IOL) power calculation and, secondly, to determine the effect of optimising the 'A' constant on the predictive accuracies of the empirical Sanders-Retzlaff-Kraff (SRK) formulas. The accuracies of the empirical formulas (SRK and SRK II) were generally similar to those of the theoretical formulas (Binkhorst II and Colenbrander-Hoffer). However the original SRK formula was more accurate than the two theoretical formulas in long eyes and in all eyes overall. Optimisation of the 'A' constants did not improve the predictive accuracy of the empirical SRK formulas.

      Register for free content

      The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.