Br J Ophthalmol 1992;76:303-306 doi:10.1136/bjo.76.5.303
  • Research Article

Calculation of the power of anterior chamber implants.

  1. J R Villada,
  2. P S Raj and
  3. T Akingbehin
  1. Department of Ophthalmology, District General Hospital, Southport.

      Abstract

      Ninety seven eyes with anterior chamber implants and a best corrected visual acuity of at least 6/12 were studied, firstly, to compare the predictive accuracies of the commonly used formulas for intraocular lens (IOL) power calculation and, secondly, to determine the effect of optimising the 'A' constant on the predictive accuracies of the empirical Sanders-Retzlaff-Kraff (SRK) formulas. The accuracies of the empirical formulas (SRK and SRK II) were generally similar to those of the theoretical formulas (Binkhorst II and Colenbrander-Hoffer). However the original SRK formula was more accurate than the two theoretical formulas in long eyes and in all eyes overall. Optimisation of the 'A' constants did not improve the predictive accuracy of the empirical SRK formulas.

      This Article

      Services

      1. Request permissions

      Responses

      1. Submit a response
      2. No responses published

      Social bookmarking

      Register for free content


      Free sample
      This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of BJO.
      View free sample issue >>

      Free archive
      The full back archive is now available for BJO. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
      Register to access the free archive >>

      Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.