Br J Ophthalmol 1998;82:252-259 doi:10.1136/bjo.82.3.252
  • Original Article
    • Clinical science

Clinical evaluation of scanning laser polarimetry: I Intraoperator reproducibility and design of a blood vessel removal algorithm

  1. Andrew Waldock,
  2. Michael J Potts,
  3. John M Sparrow,
  4. Wojciech S S Karwatowski
  1. Bristol Eye Hospital, Lower Maudlin Street, Bristol BS1 2LX
  1. Mr Andrew Waldock, Torbay Hospital, Lawes Bridge, Torquay, Devon, TQ2 7AA.
  • Accepted 10 October 1997

Abstract

AIMS To evaluate the reproducibility of the retardation values (change in polarisation) obtained with the scanning laser polarimeter in a series of normal subjects and glaucoma patients. To improve the analysis of the raw data by devising and evaluating a blood vessel removal algorithm.

METHODS Scanning laser polarimetry was performed on 10 normal subjects and 10 glaucoma patients. A series of six images was obtained from each eye. The normal subjects were re-imaged 3 months after their initial assessment. The retardation values obtained from each eye were analysed using the authors’ own methods, including the use of an algorithm to remove blood vessels from the polar profiles. The reproducibility of these measurements and the performance of the blood vessel removal algorithm were assessed.

RESULTS The “individual point” coefficient of variation was approximately 12.5% for normal subjects and 17.0% for glaucoma patients. The “integral” coefficient of variation for these groups was approximately 5.5% and 9.5% respectively. The reproducibility of the measurements did not improve with an increased number of measurements. There was no difference in the reproducibility of the measurements in normal subjects over time. The blood vessel removal algorithm improved the reproducibility of the measurements when the shape of the profile was assessed.

CONCLUSION The intraoperator reproducibility of retardation values obtained with the scanning laser polarimeter is satisfactory for its use as a clinical tool. The use of a blood vessel removal algorithm improves the reproducibility of the measurements and also assists the clinician in the interpretation of the polar profiles. Furthermore, it allows the construction of normal database polar profiles, thereby enabling the identification, location and quantification of retinal nerve fibre layer damage in an “at risk” individual’s polar profile.

Footnotes

    Register for free content


    Free sample
    This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of BJO.
    View free sample issue >>

    Free archive
    The full back archive is now available for BJO. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
    Register to access the free archive >>

    Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.