Clinical evaluation of scanning laser polarimetry: I Intraoperator reproducibility and design of a blood vessel removal algorithm

Table 5

Blood vessel removal algorithm performance (profile shape analysis)

Normal Glaucoma
Mean deviation I-Q range Mean deviation I-Q range
3 images with no algorithm 1.85 1.68–1.86 1.72 1.56–1.94
3 images with algorithm 0.56 0.52–0.58 0.62 0.56–0.67
  • I-Q range = 25th quartile to 75th quartile mean deviations for the appropriate group.