rss
Br J Ophthalmol 82:260-266 doi:10.1136/bjo.82.3.260
  • Original Article
    • Clinical science

Clinical evaluation of scanning laser polarimetry: II Polar profile shape analysis

Table 5

Correlation of retardation values and profile shape values to cup:disc ratio and visual field mean deviation

Cup:disc ratio Mean deviation (visual field)
Retardation values:
Normal +0.07; p >0.1 (0.5%) −0.03; p >0.1
Glaucoma (original) −0.08; p >0.1 (0.6%) +0.15; p >0.1
Glaucoma (standardised) −0.40; p <0.01 (16%) +0.44; p <0.01
Profile shape values:
Normal +0.01; p >0.1 (0.0%) −0.05; p >0.1
Glaucoma (original) −0.38; p <0.01 (14%) +0.39; p <0.01
Glaucoma (standardised) −0.35; p <0.01 (13%) +0.35; p <0.01
  • Statistical analysis used for the above correlations:

  • Correlation with cup:disc ratio = Pearson’s product moment correlation with Student’s t test for statistical significance and the coefficient of determination (r 2) expressed in percentages and in parentheses.

  • Correlation with visual field = Spearman’s rank correlation coefficient with statistical significance.