Table 2

Comparison of nanosecond laser and picosecond laser capsulotomy

Capsulotomy Nd:YAG ns laser Nd:YLF ps laser
Number of cases 10      9
Single pulse energy (mJ)  2.05  (1.8–2.5)      0.25
Total number of pulses 29 (11–68) 70 700 (14 800–200 000)
Total energy (mJ) 59 (22–129) 15 155 (3700–50 000)
Complications:
 IOL damage punctual pits linear pits
 increase in IOP  1      2
 corneal lesion  0      0
 repeat treatment necessary  0      2