Br J Ophthalmol 82:884-891 doi:10.1136/bjo.82.8.884
  • Original Article
    • Clinical science

Reproducibility of volumetric measurements of normal maculae with the Heidelberg retina tomograph

Table 4

Mean of three examinations of one scan compared with the mean of three examinations of another scan. The 95% CI for the mean difference of log transformed values were initially calculated and the antilogs are quoted as the 95% CI of the geometric mean ratio of the volumes of scans 1 and 2, 1 and 3, and 2 and 3

Scan 1 v scan 2 Scan 1 v scan 3 Scan 2 v scan 3
Mean difference of log values 0.0168 0.0086 −0.0082
Standard error of the differences of log values 0.0302 0.0267 0.0311
95% CI for the mean difference of log values −0.0463  to  0.08 −0.0474  to  0.0646 −0.0734  to  0.0569
95% CI for the geometric mean scan ratio 0.89 to 1.20 0.89 to 1.16 0.84 to 1.13