Br J Ophthalmol 83:157-161 doi:10.1136/bjo.83.2.157
  • Original Article
    • Clinical science

Screening for refractive errors in children: accuracy of the hand held refractor Retinomax to screen for astigmatism

Table 5

Performances of the non-cycloplegic screening for three different thresholds of manifest astigmatism (true positive case if cycloplegic cylinder value by retinoscopy or on table refractor ⩾2 D)

Manifest cylinder Positive test n (%) Negative test
n (%)
Sensitivity (%) Specificity (%) PPV (%) NPV (%)
⩾2 D
 Right eye 20 (23) 68 (77) 52 95 85 76
 Left eye 25 (29) 62 (71) 65 90 76 87
⩾1.75 D
 Right eye 23 (26) 65 (74) 58 93 83 78
 Left eye 35 (40) 52 (60) 83 81 69 90
⩾1.5 D
 Right eye 31 (35) 57 (65) 82 93 87 89
 Left eye 40 (46) 47 (54) 89 76 65 94
  • PPV = positive predictive value.

  • NPV = negative predictive value.