Br J Ophthalmol 83:276-279 doi:10.1136/bjo.83.3.276
  • Original Article
    • Clinical science

Reproducibility of fundus autofluorescence measurements obtained using a confocal scanning laser ophthalmoscope

Table 5

Greatest differences in fundus autofluorescence measurements between observers for patients (largest− smallest absolute value)

Patient HAF investigator 1 − LAF investigator 2 (arbitrary units) HAF investigator 2 − LAF investigator 1 (arbitrary units) Greatest interobserver variability (arbitrary units)
1 2.03 7.50 7.50
2 24.23 40.79 40.79
3 2.65 7.11 7.11
4 13.92 1.40 13.92
5 4.23 15.5 15.50
6 7.13 1.84 7.13
7 6.10 9.10 9.10
8 13.98 6.32 13.98
9 9.69 5.01 9.69
10 3.57 12.55 12.55
Median 6.61 7.30 11.12
  • HAF investigator 1 − LAF investigator 2 = highest value of fundus autofluorescence obtained by investigator 1 minus lowest value of fundus autofluorescence obtained by investigator 2.

  • HAF investigator 2 − LAF investigator 1 = highest value of fundus autofluorescence obtained by investigator 2 minus lowest value of fundus autofluorescence obtained by investigator 1.