A new digital optic disc stereo camera: intraobserver and interobserver repeatability of optic disc measurements

  1. G N Shuttleworth,
  2. C H Khong,
  3. J P Diamond
  1. Bristol Eye Hospital, Lower Maudlin Street, Bristol BS1 2LX
  1. Dr Shuttleworth
  • Accepted 8 December 1999

Abstract

AIMS To investigate the intraobserver and interobserver repeatability of optic disc measurement using a new digital optic disc stereo camera.

METHODS 112 consecutive new patients presenting to a glaucoma service had dilated optic disc photography performed using a new digital stereo camera (Discam, Marcher Enterprises Ltd, Hereford). The images were analysed by two masked observers using a stereo viewer and computer simulated stereopsis. Vertical and horizontal cup:disc ratios (CDR), cup area:disc area, and cup circumference:disc circumference were computed. Intraobserver and interobserver repeatability analyses were performed. Intraclass correlation coefficients (ICC) and 95% tolerance for change (TC) were computed.

RESULTS 220 optic discs were photographed, of which 196 were suitable for analysis (10 were of poor image quality and 14 had anomalous discs). Mean age of patients was 65 years, 60 were male and 48 female. For intraobserver measurements of: horizontal CDR, ICC = 0.94, TC = 0.11 (15% of range); vertical CDR, ICC = 0.92, TC = 0.14 (16% of range); cup area:disc area, ICC = 0.95, TC = 0.10 (13% of range), and cup circumference:disc circumference, ICC = 0.95, TC = 0.09 (14% of range). For interobserver measurements of: horizontal CDR, ICC = 0.89, TC = 0.14 (19% of range); vertical CDR, ICC = 0.90, TC = 0.14 (16% of range); cup area:disc area, ICC = 0.92, TC = 0.13 (16% of range), and cup circumference:disc circumference, ICC = 0.90, TC = 0.12 (17% of range). Systematic bias between observers was within acceptable limits.

CONCLUSIONS Digital stereo disc photography and analysis provide repeatable measures of optic disc variables. The results compare favourably with ophthalmoscopic and stereophotographic methods of assessment of the optic disc.

Footnotes

    Register for free content

    The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.