Efficacy of a contact lens cleaning device and its enhancement of the performance of contact lens care products
  1. C P Ifejika,
  2. L McLaughlin-Borlace,
  3. V J Lucas,
  4. A D G Roberts,
  5. J T Walker
  1. Centre for Applied Microbiology and Research, Porton Down, Salisbury, Wilts SP4 0JG
  1. Mr Ifejika charles{at}ifejika.freeserve.co.uk

Abstract

BACKGROUND Corneal infections due to contact lens contamination are risks associated with contact lens wear. Care systems which reduce these risks are desirable.

METHODS This study evaluated a contact lens cleaning device using normal saline initially and then four contact lens solutions.

RESULTS Using saline, six out of 10 tests resulted in complete removal of challenge organisms or showed reductions to 10 cfu/ml or <10 cfu/ml. Tests upon organism using multipurpose solutions showed >6 log reductions in 2–3 minutes. With laboratory made biofilms, similar results were obtained.

CONCLUSIONS This device improves lens cleaning and enhances lens care solutions. When used correctly it should lead to significant reductions in microbial keratitis associated with inadequate contact lens hygiene.

  • contact lens
  • cleaning devices
  • lens hygiene
  • keratitis

Statistics from Altmetric.com

The incidence of microbial keratitis in contact lens wear appears to be increasing and indications are that up to 30% of corneal ulcers are associated with contact lens wear.1 Lens wearers are increasingly seen in accident and emergency departments of certain hospitals.2 A contributing factor to lens wear complication is microbial contamination, and reports suggest that up to 95% of lenses in use are contaminated.3 Physical cleaning of lenses and lens cases, helps to reduce micro-organisms before chemical disinfection.4

Contact lens wearers and eye care practitioners may rely too much upon contact lens solutions to reduce microbial contamination. Studies indicate that several multipurpose solutions and peroxide systems are not able adequately to disinfect concentrations of bacteria (for example, Staphylococcus aureus) of 107–108 cfu/ml. The studies show that bacteria can survive chemical disinfection5 and they describe resistance to hydrogen peroxide by Candida parapsilosis, a yeast associated with cutaneous infections.6

Daily cleaning of contact lenses is not always properly performed by lens wearers.7 8 Levels of microbial contamination of 108 cfu/ml and above are common in used contact lens cases.1 7 This indicates that many commercially available lens care systems may not be adequate on their own. Use of monthly disposable lenses may encourage less diligence in cleaning lenses.9 Therefore, an ideal care system needs to eliminate the qualitative test of “rub and rinse” and substitute a more reliable method for cleaning contact lenses.

Devices for cleaning lenses that were marketed in the past, were generally found to be limited when tested with normal saline alone.10 Efron et al evaluated a standing wave device and an ultrasound unit,11 with standing wave providing little if any reduction to the micro-organisms tested and ultrasound producing only minimal reduction over a 22 minute period of operation.

Materials and methods

The device is commercially available and is manufactured by a LensCare Ltd. At the heart of the system is a multidirectional, vibration generating device which is completely new to this field. It is a small battery operated device which produces incredible turbulence within the lens case, resulting in very thorough cleaning.

The contact lenses used in the tests were sterile, non-ionic low water content lenses (Medalist Polymacon, Bausch and Lomb), and high water content ionic contact lenses (Acuvue daily disposable, Johnson and Johnson). The saline was Oxysept (Allergan) sterile buffered saline. The care products used were 10:10 (Ciba Vision), Optifree (Polyquad, Alcon), Complete (Polyhexamethylene, Allergan), and SOLOCare (Polyhexanide, Ciba Vision).

TEST ORGANISMS AND CULTURE MEDIA

The test organisms were: Pseudomonas aeruginosa, American Type Culture Collection (ATCC)-9027,Staphylococcus aureus (ATCC)-27735,Serratia marcescens (ATCC)-14756,Candida albicans, National Collection of Pathogenic Fungi (NCPF)-3324 and Fusarium solani, International Mycological Institute (IMI)-215765. The culture mediums used were: potato dextrose agar (PDA); tryptone soya agar (TSA), and Sabouraud dextrose agar (SDA).

EVALUATION OF THE DEVICE

The device was tested according to the repetitive treatment method detailed in EN 1174-2, 1996, “Sterilisation of medical devices, Estimation of the population of micro-organisms on product. Part 2: Guidance”. The technique describes the repeated reduction of micro-organisms by a given method until there is no significant increase in the number of micro-organisms recovered. The results accumulated from consecutive recoveries can then be evaluated. Test and control procedures were performed using normal saline.

Results

Six out of 10 tests resulted in either complete removal of challenge organisms from lenses, or reductions to 10 cfu/ml or <10 cfu/ml. Control lenses averaged only a 2.3–2.5 log reduction of test organisms (Table 1).

Table 1

Log reduction of challenge organisms using normal saline

EVALUATION WITH CONTACT LENS SOLUTIONS

The device was tested as described for saline, but substituting multipurpose solutions for saline. The treatments were performed in various cleaning cycles as illustrated in Table 2.

Table 2

The use of multipurpose solutions with the device

The results summarised in Table 2 show that a 1 minute cleaning cycle using SOLOCare (Ciba Vision) achieved a 3–4 log reduction ofS aureus from the lenses. The 2 minute cleaning cycle using Complete (Allergan) showed a 6 log reduction ofP aeruginosa from lenses, and the 3 minute cleaning cycle using SOLOCare resulted in eradication ofS marcescens.

EVALUATION WITH LENS CARE SOLUTIONS ON LABORATORY BIOFILMS

An experimental procedure developed at Moorfields Eye Hospital was used to produce simulated biofilms. The device was tested as described for saline but substituting 10:10 hydrogen peroxide (Ciba Vision) or Optifree (Alcon) for saline. The treatments were performed in two cleaning cycles and the results were as follows.

The device when used with Miraflow and 10:10 solution eradicated bacterial biofilm on the lenses. After 2 minutes of cleaning there was no recovery of the test organism. The device when used with Opticlean and Optifree showed a >10 log reduction of planktonic challenge organisms.

Discussion

This lens cleaning device when used with normal saline alone met the criteria given in the International Standards Organisation's directive ISO/CD14729.3, which gives performance tests using bacteria and fungi. Previous studies have shown that physical rubbing of a lens with cleaning solutions can reduce micro-organisms by 102–103 cfu/ml or more.4 11However, reports show that 32% of users do not clean their lenses at all after each use.12 The hands of a lens wearer may transfer contamination to a contact lens during cleaning and/or general lens manipulation.13 The new cleaning device minimises the use of the hands in lens maintenance while increasing cleaning efficacy. It is safe to use with soft, PMMA, and gas permeable contact lenses and does not change their surface properties.

Studies of Acanthamoeba keratitis report that contamination of a contact lens case with bacteria can support amoebae and enable them to feed.14 The cleaning device not only cleans lenses but also cleans its own storage case simultaneously, thus reducing the chance of lens case contamination.

Preliminary tests using a hydrogen peroxide care product surpassed the criteria given in ISO/CD14729.3 and showed a >10 log reduction of the challenge organisms. Incubating the lenses for 72 hours after cleaning showed minimal growth (100 ml withPseudomonas aeruginosa), but no colony forming units could be detected. Similar results were achieved when using Optifree and SOLOCare for 2–3 minutes.

This device should help to significantly reduce contact lens wear complications that are associated with inadequate contact lens care procedures or contaminated contact lens cases.

Acknowledgments

The authors thank John K G Dart, Michael Port, and David Shannon for their assistance in the preparation of the manuscript.

The authors also thank Shannon & Carton and Dolland & Aitchison for the supply of contact lenses.

C P Ifejika has an interest in the company LensCare Ltd.