Br J Ophthalmol 2001;85:1303-1304 doi:10.1136/bjo.85.11.1303
  • Scientific correspondence

Clinical comparison of the Keeler Pulsair 3000 with Goldmann applanation tonometry

  1. V A Parker,
  2. J Herrtage,
  3. N J C Sarkies
  1. Department of Ophthalmology, Addenbrooke's Hospital Trust, Cambridge, UK
  1. Mrs Parker
  • Accepted 13 June 2001

Abstract

AIM To confirm the accuracy of the Pulsair 3000 before introducing the instrument into clinical practice.

METHOD A masked study by two experienced tonometrists comparing the mean of the Goldmann intraocular pressure (IOP) readings against the Pulsair 3000 reading (average of four puffs). Results of 150 eyes were compared with an IOP range of 10 mm Hg–44 mm Hg.

RESULTS Correlation between the two Goldmann tonometry results was 0.9830 with a standard deviation of 1.1085 mm Hg. Correlation between the mean of the two Goldmann readings against the Pulsair 3000 reading was 0.982 with a standard deviation of 1.1179 mm Hg. Bland-Altman analysis confirms a satisfactory outcome.

CONCLUSION The Pulsair 3000 provides an accurate and objective method of measuring IOP with many advantages over traditional Goldmann tonometry.

Footnotes

    Register for free content

    The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.