Results of colour contrast sensitivity function testing without and with glare in a refractive multifocal IOL (Allergan SA 40N) through the distance focus (n = 15), through the near focus (n = 15) and in a monofocal intraocular lens (n = 11, Allergan SI 40 NB). (cpd = cycles per degree)
| cpd | SA40N distance focus | SA40N near focus | SI40NB |
| CCS function without glare in refractive multifocal IOLs and monofocal IOLs | |||
| 0.5 | 14.59 (12.02) | 15.09 (13.12) | 21.37 (11.42) |
| 1 | 9.45 (5.82) | 10.15 (7.10) | 16.20 (5.92) |
| 3 | 4.24 (2.10) | 4.22 (1.94) | 6.76 (4.23) |
| 6 | 2.97 (2.14) | 2.26 (1.08) | 3.63 (2.05) |
| 11.4 | 0.79 (0.75) | 0.95 (0.81) | 1.55 (0.94) |
| 22.8 | 0.27 (0.50) | 0.18 (0.41) | 1.11 (0.38) |
| CCS function with glare in refractive multifocal IOLs and monofocal IOLs | |||
| 0.5 | 10.37 (3.75) | 9.55 (2.29) | 13.79 (5.22) |
| 1 | 7.68 (2.64) | 8.04 (3.79) | 11.02 (3.29) |
| 3 | 3.66 (1.34) | 3.06 (2.01) | 4.50 (1.87) |
| 6 | 2.02 (1.08) | 0.94 (0.74) | 2.19 (0.69) |
| 11.4 | 0.39 (0.57) | 0.29 (0.52) | 1.08 (0.79) |
| 22.8 | 0.09 (0.30) | 0.17 (0.40) | 0.09 (0.30) |