rss
Br J Ophthalmol 2001;85:1034-1037 doi:10.1136/bjo.85.9.1034
  • Scientific correspondence

Influence of glistenings on the optical quality of acrylic foldable intraocular lens

Abstract

AIMS To assess the influence of glistenings on the optical quality of acrylic foldable intraocular lens.

METHODS Several degrees of glistenings in the optic were experimentally created by immersing the lens in water at 37°C for 48 hours and then at 25°C for 24 hours. Optical bench tests were carried out in water including measurements of spectral transmittance with the spectrophotometer, intensity of forward scattering using the integrating sphere photometer, modulation transfer function, and resolving power at various contrasts with and without the veiling glare light source.

RESULTS Glistenings of 1+ to 4+ degrees were created, among which the 4+ glistenings seemed to be extremely intense and thought to be beyond the range of clinical settings. Clinically feasible level of glistenings, up to 3+, did not adversely influence spectral transmittance, scattering, modulation transfer function, and resolving power at various contrasts. The 4+ glistenings caused mild to moderate deteriorations of the optical quality of the lens.

CONCLUSION The optical quality of the acrylic foldable intraocular lens is not significantly affected by the level of glistenings usually seen in the clinical setting.

Footnotes

    Register for free content

    The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

    Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.