rss
Br J Ophthalmol 86:394-396 doi:10.1136/bjo.86.4.394
  • Scientific correspondence

Complications of motility peg placement for porous hydroxyapatite orbital implants

Table 3

Problems associated with peg systems (48 patients with hydroxyapatite implant)

Problem No of patients %
Discharge (22 (45.8%))
    Minor/recurrent 19 39.6
    Major 3 6.2
Peg falling out (10 (20.8%))
    With overgrowth of hole 5 10.4
    With hole remaining open 5 10.4
Pyogenic granulomas (8 (16.7%))
    Around peg hole 8 16.7
    Within hole 0
Popping peg 7 14.6
Hydroxyapatite visible around peg hole 4 8.3
Part of peg shaft visible 3 6.2
Peg drilled off centre 3 6.2
Peg drilled at an angle 2 4.2
Excess movement of peg 2 4.2