In vivo confocal microscopy of the human cornea

Table 1

Intrasession coefficients of repeatability (CORs) of in vivo methods for the measurement of central corneal thickness in humans

Total cornea Epithelium
Instrument Mean (SD) COR Mean (SD) COR Ref
TSCM = tandem scanning confocal microscope; SSCM = slit scanning confocal microscope; OCT = optical coherence tomography. *Indicates same day, non-consecutive repeats (average delay of 302 (SD 83) minutes). †Indicates repeats performed on different day but at the same time of day.
Optical 539 (33) 26.6* 175
Modified optical 529 (31) 4.5 53 (9) 3.8 56
Ultrasound 581 (34) 10.1 59
Specular 533 (48) 11.8 176
Orbscan 596 (40) 11.5* 175
TSCM 532 (19) 24.5 51 (4) 7.8 55
SSCM 538 (40) 113.6 60 (13) 37.3 56
OCT 530 (32) 9.9 58 (3) 6.1† 59, 60