Br J Ophthalmol 87:960-963 doi:10.1136/bjo.87.8.960
  • Clinical science
    • Scientific reports

IOLMaster biometry: refractive results of 100 consecutive cases

Table 2

IOLMaster refractive results: comparison with ultrasound biometry (Wilcoxon signed rank test)

Prediction error Eyes within
Mean prediction error Abs prediction error Range 0.50 D 1.00 D 1.50 D 2.00 D 2.50 D
IOLMaster = IOLMaster biometry without optimisation; IOLMaster Opt = IOLMaster biometry with optimisation of the SF; US biometry Opt = ultrasound biometry with optimisation of the SF.
*IOLMaster v IOLMaster Opt, p=0.0149.
†IOLMaster Opt v US biometry Opt, p<0.0001.
‡IOLMaster v US biometry Opt, p<0.0001.
IOLMaster −0.15 (0.38) 0.29 (0.27)* 0.01 to 1.28 84% 96% 100%
IOLMaster Opt 0.00 (0.37) 0.25 (0.27)† 0.00 to 1.46 90% 96% 100%
US biometry Opt 0.00 (0.56) 0.41 (0.38)‡ 0.01 to 2.44 74% 93% 97% 99% 100%