Br J Ophthalmol 2004;88:501-504 doi:10.1136/bjo.2003.030650
  • Clinical science
    • Scientific reports

Ophthalmodynamometric determination of the central retinal vessel collapse pressure correlated with systemic blood pressure

  1. J B Jonas
  1. Correspondence to: Dr J Jonas Universitäts-Augenklinik, Theodor-Kutzer-Ufer 1-3, 68167 Mannheim, Germany; Jost.Jonasaugen.ma.uni-heidelberg.de
  • Accepted 1 September 2003

Abstract

Aims: To evaluate whether determination of the central retinal artery and vein collapse pressure correlate with systemic blood pressure measurements, using a new Goldmann contact lens associated ophthalmodynamometric device

Methods: The prospective clinical study included 92 eyes of 92 patients presenting with cataract or refractive problems (n = 40; control study group) or with retinal and orbital pathologies (n = 52). With topical anaesthesia, a Goldmann contact lens fitted with a pressure sensor in its holding ring was placed onto the cornea. Pressure was asserted onto the globe by pressing the contact lens, and the pressure value at the time when the central retinal artery and vein started pulsating were noted as central retinal artery and vein collapse pressure. Additionally, the brachial arterial blood pressure was measured.

Results: In the control study group, central retinal artery collapse pressure was highly significantly correlated with diastolic blood pressure (correlation coefficient r = 0.77; p<0.001) and systolic blood pressure (r = 0.35; p = 0.03). Central retinal vein collapse pressure was statistically independent of diastolic blood pressure (p = 0.11). In eyes with retinal or orbital diseases, the correlation coefficients were lower than in the control study group. In eyes with retinal arterial occlusions, central retinal vessel collapse pressure measurements were not correlated with arterial blood pressure measurements.

Conclusions: Depending on coexisting retinal or orbital diseases, ophthalmodynamometric estimation of the central retinal artery collapse pressure, performed during a routine Goldmann contact lens ophthalmoscopy, correlates with systemic blood pressure measurements.

Footnotes

  • Proprietary interest: none.

This Article

Services

  1. Request permissions

Responses

  1. Submit a response
  2. No responses published

Social bookmarking

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of BJO.
View free sample issue >>

Free archive
The full back archive is now available for BJO. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.