Br J Ophthalmol 2004;88:1030-1035 doi:10.1136/bjo.2003.036699
  • Clinical science
    • Extended reports

A newly developed peripheral anterior chamber depth analysis system: principle, accuracy, and reproducibility

  1. K Kashiwagi,
  2. F Kashiwagi,
  3. Y Toda,
  4. K Osada,
  5. T Tsumura,
  6. S Tsukahara
  1. Department of Ophthalmology, University of Yamanashi Faculty of Medicine, Tamaho Yamanashi, Japan
  1. Correspondence to: Kenji Kashiwagi MD Department of Ophthalmology, University of Yamanashi Faculty of Medicine, 1110 Shimokato, Tamaho, Yamanashi 409-3898, Japan; kenjikyamanashi.ac.jp
  • Accepted 7 December 2003

Abstract

Aim: To develop a new, non-contact system for measuring anterior chamber depth (ACD) quantitatively, and to investigate its accuracy as well as interobserver and intraobserver reproducibility.

Methods: The system scanned the ACD from the optical axis to the limbus in approximately 0.5 second and took 21 consecutive slit lamp images at 0.4 mm intervals. A computer installed program automatically evaluated the ACD, central corneal thickness (CT), and corneal radius of curvature (CRC) instantly. A dummy eye was used for investigating measurement accuracy. The effects of CT and CRC on the measurement results were examined using a computer simulation model to minimise measurement errors. Three examiners measured the ACD in 10 normal eyes, and interobserver and intraobserver reproducibility was analysed.

Results: The ACD values measured by this system were very similar to theoretical values. Increase of CRC and decrease in CT decreased ACD and vice versa. Data calibration using evaluated CT and CRC successfully reduced measurement errors. Intraobserver and interobserver variations were small. Their coefficient variation values were 7.4% (SD 2.3%) and 6.7% (0.7%), and these values tended to increase along the distance from the optical axis.

Conclusion: The current system can measure ACD with high accuracy as well as high intraobserver and interobserver reproducibility. It has potential use in measuring ACD quantitatively and screening subjects with narrow angle.

Footnotes

  • Proprietary interest: Japanese patent application No 2003-111322.

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.