Br J Ophthalmol 2005;89:1474-1477 doi:10.1136/bjo.2005.071738
  • Clinical science
    • Extended reports

Linear deposits on the surfaces of intraocular lenses implanted through a hexagonal cartridge which mimic scratches/cracks on the lenses

  1. G Kleinmann1,
  2. A L Marcovich2,
  3. D J Apple1,
  4. N Mamalis1
  1. 1David J Apple, MD, Laboratories for Ophthalmic Devices Research, John A Moran Eye Center, Department of Ophthalmology and Visual Sciences, University of Utah, Salt Lake City, UT, USA
  2. 2Department of Ophthalmology, Kaplan Medical Center, Rehovot, Israel
  1. Correspondence to: Guy Kleinmann MD, John A Moran Eye Center, Department of Ophthalmology and Visual Sciences, University of Utah, 50 N Medical Drive, Salt Lake City, UT 84132, USA; guy.kleinmannhsc.utah.edu
  • Accepted 1 June 2005

Abstract

Aim: To describe unique linear deposits on the surface of posterior chamber intraocular lenses (IOLs) occurring after implantation through a hexagonal cartridge.

Methods: Five ACR6D SE IOLs (Corneal Laboratories, Pringy, France) were injected/implanted through hexagonal cartridges. Two of these were injected into a petri dish and the remaining three were inserted into the evacuated capsular bag of cadaver eyes. In addition, three other ACR6D SE IOLs were injected into a petri dish through round cartridges. The latter served as controls.

Results: All lenses that were injected/implanted through the hexagonal cartridges demonstrated linear deposits on the posterior surface of the IOL optic component. The IOLs that were injected through the round cartridges had no deposits. All the hexagonal cartridges showed signs of internal cracking. None of the round cartridges were cracked.

Conclusion: Implantation of at least one IOL model, the ACR6D SE IOL, through a hexagonal cartridge can result in linear deposits on the posterior optical surface of the IOL. The shape of the cartridge appears to be a significant factor in causing the depositions.

Footnotes

  • The authors have no financial or proprietary interest in any product mentioned in this paper.

  • Video presentation at the Annual Meeting of the American Society of Cataract and Refractive Surgery, Washington DC, USA, April 2005.

This Article

Services

  1. Request permissions

Responses

  1. Submit a response
  2. No responses published

Social bookmarking

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of BJO.
View free sample issue >>

Free archive
The full back archive is now available for BJO. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.