Aniseikonia associated with epiretinal membranes
  1. M Ugarte,
  2. T H Williamson
  1. Department of Ophthalmology, Queen Mary’s Hospital, Sidcup, Kent DA14 6LT and St Thomas’s Hospital, London SE1 7EH, UK
  1. Correspondence to: Marta Ugarte Department of Ophthalmology, St Thomas’s Hospital, London SE1 7EH, UK; mugartedoctors.org.uk

Abstract

Aims: To determine whether the computerised version of the new aniseikonia test (NAT) is a valid, reliable method to measure aniseikonia and establish whether aniseikonia occurs in patients with epiretinal membranes (ERM) with preserved good visual acuity.

Methods: With a computerised version of the NAT, horizontal and vertical aniseikonia was measured in 16 individuals (mean 47 (SD 16.46) years) with no ocular history and 14 patients (mean 67.7 (14.36) years) with ERM. Test validity was evaluated by inducing aniseikonia with size lenses. Test reliability was assessed by the test-retest method.

Results: In normal individuals, the mean percentage (SD) aniseikonia was −0.24% (0.71) horizontal and 0% (0.59) vertical. Validity studies revealed mean (SD) 0.990 (0.005) horizontal and 0.991 (0.004) vertical correlation coefficients, 0.985 (0.111) horizontal and 0.989 (0.102) vertical slope. Repeatability coefficients were 1.04 horizontal and 0.88 vertical. Aniseikonia in patients with ERM ranged from 4% to 14%. Eight patients showed 2% or more size difference between horizontal and vertical meridians.

Conclusions: The aniseikonia test used in this study can be considered a simple, fast, valid and reliable method to measure the difference in image size perceived by each eye. Aniseikonia does occur in symptomatic patients with ERM. The effect of ERM on image size is heterogeneous across the retinal area affected.

  • ERM, epiretinal membranes
  • NAT, new aniseikonia test
  • VA, visual acuity
  • micropsia
  • macropsia
  • aniseikonia
  • epiretinal membranes
  • ERM, epiretinal membranes
  • NAT, new aniseikonia test