Br J Ophthalmol 2005;89:1609-1615 doi:10.1136/bjo.2005.070409
  • Clinical science
    • Extended reports

Lower limits of fluorescein and indocyanine green dye for digital cSLO fluorescence angiography

  1. A Bindewald,
  2. O Stuhrmann,
  3. F Roth,
  4. S Schmitz-Valckenberg,
  5. H-M Helb,
  6. A Wegener,
  7. N Eter,
  8. F G Holz
  1. Department of Ophthalmology, University of Bonn, Ernst-Abbe-Strasse 2, D-53127 Bonn, Germany
  1. Correspondence to: Frank G Holz MD, Department of Ophthalmology, University of Bonn, Ernst-Abbe-Strasse 2, D-53127 Bonn, Germany; frank.holzukb.uni-bonn.de
  • Accepted 1 July 2005

Abstract

Background: With the advent of digital confocal scanning laser ophthalmoscopy it is possible to detect low levels of fluorescence. Here we used a novel confocal scanning laser ophthalmoscope (cSLO) to determine lower limits of dye required for fluorescein (FL) and indocyanine green (ICG) angiography.

Methods: A cSLO (Heidelberg retina angiograph 2, Heidelberg Engineering, Dossenheim, Germany) with an optically pumped solid state laser (488 nm) for FL and a diode laser (790 nm) for ICG angiography (FL/ICG-A) was used. 62 FL-As were performed in 53 patients and 45 ICG-As were performed in 39 patients with neovascular age related macular degeneration. The volume and overall dye content of bolus injections was gradually tapered (FL: 500 mg, 250 mg, 200 mg, 166 mg, 100 mg; ICG: 25 mg, 20 mg, 15 mg, 10 mg, 5 mg, 2.5 mg), while dye concentrations were kept constant at 100 mg/ml for FL and at 5 mg/ml for ICG. Images were obtained 1, 5, 15, and 30 minutes after dye injection. Image quality was evaluated by two independent readers using standardised criteria.

Results: For amounts down to 166 mg for FL and to 5 mg for ICG, sufficient image quality was achieved during all phases following injection. Only late phase images showed less contrast compared to typically used dye amounts, which was irrelevant for interpretation and clinical management.

Conclusions: With the increased sensitivity of this novel cSLO system, amounts of injected dye during FL-A can be reduced to one third for FL and to one fifth for ICG without relevant loss of image quality or information compared to conventionally used dye levels. These amounts can be used for routine angiography and allow relevant savings for units performing FL-A.

Footnotes

  • Competing interests: none declared

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of BJO.
View free sample issue >>

Free archive
The full back archive is now available for BJO. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.