Br J Ophthalmol 2006;90:1519-1523 doi:10.1136/bjo.2006.097766
  • Clinical science
    • Extended reports

Investigation of the stability of one-piece acrylic intraocular lenses in cataract surgery and in combined vitrectomy surgery

  1. T Iwase1,
  2. K Sugiyama2
  1. 1Toyama Prefectural Central Hospital, Toyama, Japan
  2. 2Kanazawa University Graduate School of Medical Science, Kanazawa, Japan
  1. Correspondence to: T Iwase Department of Ophthalmology, Toyama Prefectural Central Hospital, 2–2–78 Nishinagae, Toyama City, Toyama 930–8550, Japan; TsuyoshiIwase{at}aol.com
  • Accepted 6 August 2006
  • Published Online First 16 August 2006

Abstract

Background: To compare the degrees of intraocular lens (IOL) movement between eyes that received a one-piece acrylic IOL and those that received a three-piece acrylic IOL after cataract surgery, and also among eyes that received a one-piece acrylic IOL after a combined vitrectomy surgery for cataract.

Methods: In the first study, we report on 50 patients who were implanted with a one-piece acrylic IOL in one eye and a three-piece acrylic IOL in the contralateral eye for senile cataract. In the second study, we report on 50 patients who were implanted with a one-piece acrylic IOL in combined vitrectomy surgery for cataract and retinal diseases. The degree of IOL decentration and tilt, and anterior chamber depth (ACD) were measured using Scheimpflug video photography at 1 week, 1 month, 3 months and 6 months after surgery in both the studies. The postoperative refractive status was also examined.

Results: The mean decentration and tilt showed no marked changes during the follow-up in eyes with either IOL implanted, and no marked differences were noted in either study throughout the follow-up. The ACD did not change after surgery with one-piece IOL implantation, for either the cataract surgery group or the combined surgery group, except for 1 week after surgery in eyes requiring gas tamponade. In contrast, marked shallowing in ACD was observed in the three-piece group after cataract surgery. The spherical equivalent did not change markedly in either study.

Conclusions: The one-piece acrylic IOL was stable in the capsular bag, both horizontally and vertically, after cataract surgery, and also after combined surgery.

Footnotes

  • Published Online First 16 August 2006

  • Competing interests: None.

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of BJO.
View free sample issue >>

Free archive
The full back archive is now available for BJO. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.