Background/aims: The effectiveness of a self administered eye drop medication can only be assessed if the compliance is known. The authors studied the specificity and sensitivity of a new microprocessor controlled monitoring device.

Methods: The monitoring system was conducted by an 8 bit microcontroller for data acquisition and storage with sensors measuring applied pressure to the bottle, temperature, and vertical position. 10 devices were mounted under commercial 10 ml eye drops. Test subjects had to note down each application manually. A total of 15 applications each within 3 days was intended.

Results: Manual reports confirmed 15 applications for each of the 10 bottles. The monitoring devices detected a total of 149 events; one was missed; comprising a sensitivity of 99%. Two devices registered three applications, which did not appear in the manual protocols, indicating a specificity of about 98%. Refrigerated bottles were correctly identified. The battery lifetime exceeded 60 days.

Conclusion: The new monitoring device demonstrated a high reliability of the collected compliance data. The important, yet often unknown, influence of compliance in patient care and clinical trials shall be illuminated by the new device. This may lead to a better adapted patient care. Studies will profit from a higher credibility and results will be less influenced by non-compliance.

Footnotes

  • Sponsors: none.

  • Competing interests: Both authors have a proprietary interest in the monitoring unit.

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.