Br J Ophthalmol 2008;92:1377-1381 doi:10.1136/bjo.2007.134080
  • Original Article
    • Clinical science

Retinal nerve fibre layer assessment in myopic glaucomatous eyes: comparison of GDx variable corneal compensation with GDx enhanced corneal compensation

This article has been Unlocked
  1. S Morishita,
  2. T Tanabe,
  3. S Yu,
  4. M Hangai,
  5. T Ojima,
  6. H Aikawa,
  7. N Yoshimura
  1. Department of Ophthalmology and Visual Sciences, Kyoto University Graduate School of Medicine, Kyoto, Japan
  1. Dr T Tanabe, Department of Ophthalmology, The Tazuke Kofukai Medical Institute, Kitano Hospital, 2-4-20 Ohugimachi, Kita-Ku, Osaka 530-8480, Japan; t-tanabe{at}kitano-hp.or.jp
  • Accepted 6 July 2008
  • Published Online First 29 July 2008

Abstract

Aim: To compare the results of scanning laser polarimetry (GDx) with variable corneal compensation (VCC) and enhanced corneal compensation (ECC) when applied to myopic glaucomatous eyes.

Methods: Forty glaucoma eyes with moderate myopia (between −3 and −6 D) and 35 glaucoma eyes with high myopia (−8 D or greater) were enrolled in this study. GDx VCC, GDx ECC and standard automated perimetry (SAP) were performed. The prevalence of an atypical retardation pattern (ARP), the typical scan score (TSS) and retinal nerve fibre layer (RNFL) thickness were compared between VCC and ECC in both groups of myopic subjects. A correlation analysis between RNFL thickness and visual sensitivity was also conducted.

Results: In both myopic groups, the mean TSS is significantly lower (p<0.0001), and the prevalence of ARP was significantly higher (p<0.0001) by VCC scans than by ECC scans. Temporal, superior, nasal, inferior, temporal (TSNIT) average and temporal average thickness showed significantly higher values (p<0.001) by VCC than by ECC. A statistically significant association was observed between TSNIT average and mean deviation of SAP by ECC scan.

Conclusions: ECC scans showed a better retardation pattern and structure–function relationship than did VCC, and ECC appeared to be more suitable for RNFL assessment in glaucomatous eyes that are moderately to highly myopic.

Footnotes

  • Competing interests: None.

  • Ethics approval: The Ethics Committee of Kyoto University Graduate School of Medicine approved all protocols.

  • Patient consent: Obtained.

  • This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

This article has been Unlocked

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.