Br J Ophthalmol 2008;92:1612-1616 doi:10.1136/bjo.2007.129932
  • Original Article
    • Clinical science

Quantitative analysis of anterior segment optical coherence tomography images: the Zhongshan Angle Assessment Program

  1. J W Console1,2,
  2. L M Sakata3,
  3. T Aung3,
  4. D S Friedman4,
  5. M He1
  1. 1
    Zhongshan Ophthalmic Center, Guangzhou, People’s Republic of China
  2. 2
    University of Kansas Medical School, Kansas City, Kansas, USA
  3. 3
    Singapore Eye Research Institute and Singapore National Eye Center, Singapore
  4. 4
    Wilmer Eye Institute, Johns Hopkins University, Baltimore, Maryland, USA
  1. Dr M He, Zhongshan Ophthalmic Center, Guangzhou 510060, People’s Republic of China; mingguang_he{at}yahoo.com
  • Accepted 3 June 2008
  • Published Online First 10 July 2008

Abstract

Aim: To describe a new image analysis method and software for anterior chamber images obtained by the anterior-segment optical coherence tomography (AS-OCT) and to assess its intraobserver and interobserver measurement reproducibility.

Methods: Twenty 8-bit greyscale 600×300 AS-OCT images with apparent wide angles and 20 images with apparent narrow angles were consecutively selected from a database. Two glaucoma fellowship-trained ophthalmologists used proprietary image analysis software to analyse the images twice. Algorithms defined the borders and curvatures of anterior chamber (AC) structures and measured AC parameters using scleral spur location as the only observer input. The intraobserver and interobserver reproducibility of scleral spur location and angle parameters was calculated in terms of limits of agreement (LOA; mean of differences±1.96SD of differences) and coefficient of variation (CV; SD of differences/overall mean).

Results: The analysis software successfully measured all parameters in all images. When the same image was assessed twice by the same grader, the mean differences ranged from 0 to 0.010 mm in linear measurements and 0.001 to 0.006 mm2 in angle area measurements. LOA tended to be greater in the wider angles. The upper and lower limit values of LOAs were approximately 1/5 to 1/4 of the overall mean. Measurements between two graders had a higher variance. Reproducibility in terms of CV was better in wide angles when compared with narrow angles. The reproducibility of scleral spur placement between observers was poorer in narrow angles (p = 0.001). About 50% of the interobserver variance in angle-area measurements was attributable to the variance of scleral spur placement, while this proportion was only 10–20% in linear measurements.

Conclusions: Determination of angle parameters using semiautomated software leads to variability in measurement. Variability increases when more than one observer identifies the scleral spur. Variability differs in narrow and open angles, and so including both types is essential when evaluating angle-assessment software. A fully automated analysis and higher image resolution would likely improve quantification of Visante AS-OCT images.

Footnotes

  • Competing interests: DSF and TA have received instruments loaned by Zeiss-Meditec, Heidelberg Engineering, and Takagi Instruments and TA has received research assistance from Zeiss Meditec.

  • Ethics approval: Ethics approval was provided by the Institional Review Board of the Singapore Eye Research Institute.

  • Patient consent: Obtained.

Register for free content


Free sample
This recent issue is free to all users to allow everyone the opportunity to see the full scope and typical content of BJO.
View free sample issue >>

Free archive
The full back archive is now available for BJO. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006, back to volume 1 issue 1.
Register to access the free archive >>

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.