Measurement of total blood flow in the normal human retina using Doppler Fourier-domain optical coherence tomography

  1. Y Wang1,
  2. A Lu1,
  3. J Gil-Flamer1,
  4. O Tan1,
  5. J A Izatt2,
  6. D Huang1
  1. 1
    Center for Ophthalmic Optics & Lasers, Doheny Eye Institute and Department of Ophthalmology, Keck School of Medicine, University of Southern California, Los Angeles, California, USA
  2. 2
    Department of Biomedical Engineering, Duke University, Durham, North Carolina, USA
  1. Dr Y Wang, Doheny Eye Institute—DVRC 160B, 1355 San Pablo Street, Los Angeles, CA 90033, USA; yiminwan{at}usc.edu
  • Accepted 9 December 2008
  • Published Online First 23 January 2009

Abstract

Aim: To measure total retinal blood flow in normal human eyes using Doppler Fourier-domain optical coherence tomography (FD-OCT).

Methods: 10 normal people aged 35 to 69 years were measured for the right eye using Doppler FD-OCT. Double circular scans around the optic nerve heads were used. Four pairs of circular scans that transected all retinal branch vessels were completed in 2 s. Total retinal blood flow was obtained by summing the flows in the branch veins. Measurements from the eight scans were averaged. Veins with diameters >33 μm were taken into account.

Results: Total retinal blood flow could be measured in eight of 10 subjects: mean (SD) = 45.6 (3.8) μl/min (range 40.8 to 52.9 μl/min). The coefficient of variation for repeated measurements was 10.5%. Measured vein diameters ranged from 33.3 to 155.4 μm. The averaged flow speed was 19.3 (2.9) mm/s, which did not correlate with vessel diameter. There was no significant difference between flows in the superior and inferior retinal hemispheres.

Conclusions: Double circular scanning using Doppler FD-OCT is a rapid and reproducible method to measure total retinal blood flow. These flow values are within the range previously established by laser Doppler flowmetry.

Footnotes

  • Competing interests: DH receives royalty from the Massachusetts Institute of Technology derived from an OCT patent licensed to Carl Zeiss Meditec and receives stock options, travel support, research grants, and patent royalties from Optovue. JAI is a cofounder and corporate officer of Bioptigen; YW: Optovue.

  • Ethics approval: Ethics approval was provided by the institutional review board of the University of Southern California.

Register for free content


Free trial
Individuals may register for a free 60 day online trial to all content.

Free archive
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.