Light-scattering characteristics of explanted opacified Aquasense intraocular lenses
- 1Department of Ophthalmology, Academic Medical Center, Amsterdam, The Netherlands
- 2Oxford Eye Hospital, Headington, Oxford, UK
- 3Netherlands Institute for Neuroscience, Netherlands Royal Academy, Amsterdam, The Netherlands
- Dr I van der Meulen, A2 Room 123.1, Department of Ophthalmology, Academic Medical Center, Meibergdreef 9, 1100 DD Amsterdam, The Netherlands; i.j.vandermeulen{at}amc.uva.nl
- Accepted 12 November 2008
- Published Online First 5 December 2008
Abstract
Aim: To study forward light-scattering characteristics of calcified explanted intraocular lenses (IOLs) (Aquasense, Ophthalmic Innovation International Ontario).
Methods: The amount of light scattered by the opacified IOLs was measured using a validated in vitro set-up for angles from 1.7° to 22°. This set-up gives results directly comparable with straylight values as valid for the in vivo situation.
Results: Straylight is highest at large angles and declines steeply approaching 0° angle. This corresponds to the in vivo findings that opacified IOLs cause important visual complaints but have little effect on visual acuity. At 7.5°, log (s) is around 1.8 and 2.9 for the two lenses respectively. This corresponds to 8× and 100× increases in straylight values compared with values in young, normal eyes.
Conclusion: High straylight values caused by opacified IOLs can explain subjective complaints of reduced quality of vision in patients with opacified implants, despite good visual acuity.
Footnotes
-
Competing interests: None.
-
Funding: This study was supported in part by Dutch Grant IS062031.