Light scattering characteristics of explanted opacified Aquasense™ intraocular lenses
  1. Ivanka JE van der Meulen (i.j.vandermeulen{at}amc.uva.nl),
  2. Hamid Porooshani (proushan{at}gmail.com),
  3. Thomas JTP Van den Berg (t.j.vandenberg{at}nin.knaw.nl)
  1. Academic Medical Center, Netherlands
  2. Oxford Eye Hospital, United Kingdom
  3. Netherlands Institute for Neuroscience, Netherlands

    Abstract

    Aim: To study forward light scattering characteristics of calcified explanted intraocular lenses (IOLs) (Aquasense, Ophthalmic Innovation international Ontario, California, USA).

    Methods: The amount of light scattered by the opacified IOLs was measured using a validated in vitro set-up for angles from 1.7º to 22º. This set-up gives results directly comparable to straylight values as valid for the in vivo situation.

    Results: Straylight is highest at large angles, and declines steeply approaching 0º angle. This corresponds to the in vivo findings that opacified IOLs cause important visual complaints, but have little effect on visual acuity. At 7.5º log (s) is around 1.8 and 2.9 for the two lenses respectively. This corresponds to 8 x and 100 x increases in straylight values compared to values in young, normal eyes.

    Conclusion: High straylight values caused by opacified IOLs can explain subjective complaints of reduced quality of vision in patients with opacified implants, despite good visual acuity.