Table 1

The comparison of values between IOLMaster 500 and IOLMaster 700

nDeviceMeanSDPaired samples t test (p value)Range (minimum–maximum)ICCDifferences of the meansSD of difference of means95% CI of differences of meansPaired samples t test p value
Anterior Chamber depth (mm)175IOLmaster 5003.310.38<0.0012.23–4.190.920−0.080.13−0.102 to −0.063<0.001
IOLMaster 7003.230.372.33–4.14
Axial length (mm)171IOLMaster 50023.6271.250.00820.92–28.331.0−0.0050.02−0.008 to −0.0020.830
IOLMaster 70023.6221.3220.91–29.69
Keratometry (flat, D)188IOLMaster 50043.461.770.00148.35–37.170.992−0.050.22−0.079 to −0.014<0.001
IOLMaster 70043.411.7337.25–48.21
Keratometry (steep, D)188IOLMaster 50044.501.81<0.00149.13–38.140.989−0.080.27−0.114 to −0.036<0.001
IOLMaster 70044.421.8338.05–49.07
  • Difference of the means was computed by subtracting IOLMaster 500 values form IOLMaster 700 values.

  • ICC, intraclass correlation coefficient.