Table 1.2

Differences in IOP between different tonometers and Goldmann Applanation Tonometry (GAT). Pooled estimates and summary 95% limits of agreement11-45 .

TonometerMean Difference between Tonometer and GAT95% Confidence Interval95% Limits of Agreement% within 2 mmHg
DCT1.8+1.3+2.3-3.0+6.647
NCT0.3-0.1+0.7-3.5+4.069
ORA1.5+0.9+2.2-4.3+7.345
Ocuton S2.7-1.2+6.7-4.0+9.633
RT-(Icare)0.9+0.5+1.5-4.3+6.351
TonoPen0.2-0.4+0.9-5.2+5.752
Transpalpebral-0.5-1.3+0.3-7.0+5.945
  • DCT = Dynamic Contour Tonometer; NCT = Non-Contact Tonometer; ORA = Ocular Response Analyzer; RT = Rebound Tonometer.