Abstract
Fluorometric studies have been made with modified slit-lamp microscopes. These instruments offered considerable versatility in modifying experimental parameters, but were seldom optimized in terms of detector and illuminator apertures, filter band-pass characteristics, detection system, light-source stabilization, or optical resolution. Improper control of these parameters can result in artifacts which override useful information.
A new fluorophotometer for ocular research has been designed that utilizes an optical system, which eliminates the need for a contact lens. An effort was made to optimize both the axial resolution and the minimum detectable signal.
The characteristics of standard barrier filters and excitation filters were examined and compared with photomultiplier sensitivity curves. It was necessary to use double filters in order to reduce artifacts due to excitation light that was not rejected by the barrier filter.
In order to optimize the signal-to-noise ratio, a photon counting detection system was used. A dark current equivalent to 3 × 10−10 g/ml allows a useful detection range of 1 × 10−10 g/ml.
Similar content being viewed by others
References
Brubaker RF, Coakes RL (1978) Use of a xenon flash tube as the excitation source in a new slit-lamp fluorophotometer. Am J Ophthalmol 86:474
Cunha-Vaz J, Faria de Abreu JR, Campos AJ, Figo GM (1975) Early breakdown of the blood-retinal barrier in diabetes. Br J Ophthalmol 59:649–656
Delori FC, Ben-Sira I, Trenpe C (1976) Fluorescein angiography with an optimized filter combination. Am J Ophthalmol 82:559–566
Maurice DM (1963) A new objective fluorophotometer. Exp Eye Res 2:33–38
Waltman SR, Kaufman HE (1970) A new objective slit lamp fluorophotometer. Invest Ophthalmol 9:247–249
Waltman SR, Oestrich C, Krupin T, Hanish S, Ratan S, Santiago J, Kilo C (1978) Quantitative vitreous fluorophotometry. Diabetes 27: 85–87
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Munnerlyn, C.R., Gray, J.R. & Hennings, D.R. Design considerations for a fluorophotometer for ocular research. Graefe's Arch Clin Exp Ophthalmol 222, 209–211 (1985). https://doi.org/10.1007/BF02133676
Issue Date: