Optical performance of multifocal intraocular lenses

J Cataract Refract Surg. 1990 Jul;16(4):413-22. doi: 10.1016/s0886-3350(13)80793-6.

Abstract

The optical performance of one monofocal and five multifocal lenses was evaluated in the laboratory and photographically. The laboratory testing included determination of the modulation transfer function (MTF), through focus response (TFR), resolution efficiency, and Strehl ratio of each lens. The photographic testing included photographs of the Regan high contrast acuity chart at ten feet with clearest focus and 18 additional photographs in which the image was defocused using minus trial lenses in 0.25 diopter increments. A color photograph of the Kodak color chart was also taken using each lens. All testing was conducted using a 3 mm artificial pupil under ideal implant conditions with no decentration or tilt. The laboratory and photographic results demonstrate that all the multifocal lenses had a two- to three-fold increase in the depth of field with at least a 50% lower contrast in the retinal image. The photographic testing revealed a one to two line better resolution limit with the monofocal lens, which corresponded to the 12% to 41% better MTF cut-off value with the monofocal lens by laboratory testing. The measured resolution efficiencies of all six lenses were comparable. The color photographs revealed color mixing of adjacent colors with the multifocal lenses, whereas the colors appeared unchanged from the original with the monofocal lens.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Equipment Design
  • Evaluation Studies as Topic
  • Lenses, Intraocular / standards*
  • Optics and Photonics / classification*
  • Photography