Biomechanical assessment of healthy and keratoconic corneas (with/without crosslinking) using dynamic ultrahigh-speed Scheimpflug technology and the relevance of the parameter (A1L−A2L)
  1. Thomas A Fuchsluger1,2,
  2. Susanne Brettl3,4,
  3. Gerd Geerling2,
  4. Wolfgang Kaisers5,
  5. Philipp Franko Zeitz4
  1. 1 Department of Ophthalmology, Erlangen University Hospital, Friedrich-Alexander-University of Nuremberg-Erlangen, Erlangen, Germany
  2. 2 Department of Ophthalmology, Dusseldorf University Hospital, Heinrich-Heine-University Dusseldorf, Düsseldorf, Germany
  3. 3 Department of Ophthalmology, Charité University Medicine Berlin, Berlin, Germany
  4. 4 Zeitz Franko Zeitz Ophthalmology Specialists Practice, Dusseldorf, Germany
  5. 5 Department of Anesthesiology, Dusseldorf University Hospital, Heinrich-Heine-University Dusseldorf, Düsseldorf, Germany
  1. Correspondence to Professor Thomas A Fuchsluger, Department of Ophthalmology, Friedrich-Alexander University, Erlangen, Schwabachanlage 6, 91054, Germany; thomas.fuchsluger{at}uk-erlangen.de

Abstract

Aims To examine corneal biomechanics in healthy and keratoconic eyes, with or without crosslinking obtained by ultrahigh-speed Scheimpflug measurements (Corvis ST).

Methods One hundred and seventeen eyes were studied in three groups: group 1 (n=39) contained keratoconic eyes without crosslinking. Group 2 (CXL; n=28) comprised keratoconic eyes after crosslinking. These were compared with a control group (n=50 matched healthy eyes). In addition, 10 keratoconus patients, before and after CXL treatment, respectively, were examined.

Results The novel parameter A1L–A2L demonstrated highly significant differences between crosslinked corneas and untreated keratoconic or healthy corneas. Velocity during second applanation (A2V) and deformation amplitude (DA) were significantly increased in crosslinked keratoconic eyes both compared with untreated keratoconic eyes and with healthy controls. Radius at highest curvature also was significant among all groups. Inward applanation length (A1L) was significantly increased in controls, whereas outward applanation length (A2L) was significantly reduced in crosslinked keratoconic eyes compared with both other groups. The follow-up analysis revealed statistically significant changes in pachymetry and intraocular pressure and showed tendencies towards significance in applanation times 1 and 2 and in DA.

Conclusions Both A2V and A2L are viable parameters to discriminate healthy from keratoconic but also crosslinked from non-crosslinked keratoconic corneas. The difference of A1L−A2L could reliably discriminate crosslinked from non-crosslinked and healthy corneas. Follow-up examination in a small cohort allows distinction between crosslinked and untreated keratoconus in follow-up examinations. The difference of A1L-A2L could reliably discriminate crosslinked from non-crosslinked and healthy corneas. Measurements of corneal deformation using dynamic ultrahigh-speed Scheimpflug technology are reproducible and provide useful information about keratoconus assessment and biomechanics. Therefore, the Corvis ST seems to provide useful technology to monitor therapeutic success of crosslinking treatment.

  • corvis ST
  • keratoconus
  • crosslinking
  • biomechanics
  • diagnostic instrument
  • cornea

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • Contributors All persons who have made substantial contributions to the work reported in the manuscript met the criteria for authorship and are mentioned as authors and co-authors.

  • Funding The authors have not declared a specific grant for this research from any funding agency in the public, commercial or not-for-profit sectors.

  • Competing interests TAF, SB, WK and GG do not have financial, general and institutional competing interests.

    PFZ has been consulting for Oculus Optikgeraete GmbH on several congresses and meetings.

  • Patient consent Parental/guardian consent obtained.

  • Ethics approval Heinrich-Heine-University Dusseldorf Ethics Committee approval was obtained (study-ID 4628).

  • Provenance and peer review Not commissioned; externally peer reviewed.