Predicting Humphrey 10-2 visual field from 24-2 visual field in eyes with advanced glaucoma

Abstract

Aims To predict Humphrey Field Analyzer Central 10-2 Swedish Interactive Threshold Algorithm-Standard test (HFA 10-2) results (Carl Zeiss Meditec, San Leandro, CA) from HFA 24-2 results of the same eyes with advanced glaucoma.

Methods Training and testing HFA 24-2 and 10-2 data sets, respectively, consisted of 175 eyes (175 patients) and 44 eyes (44 patients) with open advanced glaucoma (mean deviation of HFA 24-2 ≤−20 dB). Using the training data set, the 68 total deviation (TD) values of the HFA 10-2 test points were predicted from those of the innermost 16 HFA 24-2 test points in the same eye, using image processing or various machine learning methods including bilinear interpolation (IP) as a standard for comparison. The absolute prediction error (PredError) was calculated by applying each method to the testing data set.

Results The mean (SD) test–retest variability of the HFA 10-2 results in the testing data set was 2.1±1.0 dB, while the IP method yielded a PredError of 5.0±1.7 dB. Among the methods tested, support vector regression (SVR) provided a smallest PredError (4.0±1.5 dB). SVR predicted retinal sensitivity at HFA 10-2 test points in the preserved ‘central isle’ of advanced glaucoma from HFA 24-2 results of the same eye within an error range of about 25%, while error range was approximately twice of the test–retest variability.

Conclusion Applying SVR to HFA 24-2 results allowed us to predict TD values at HFA 10-2 test points of the same eye with advanced glaucoma with an error range of about 25%.

  • glaucoma
  • field of vision

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.