Variation of nerve fibre layer thickness measurements with age and ethnicity by scanning laser polarimetry
  1. D Poinoosawmya,
  2. L Fontanaa,b,
  3. J X Wua,
  4. F W Fitzkeb,
  5. R A Hitchingsa
  1. aGlaucoma Unit, Moorfields Eye Hospital, London, bInstitute of Ophthalmology, University College London, London
  1. Mr Roger A Hitchings, Glaucoma Unit, Moorfields Eye Hospital, City Road, London EC1V 2PD.

Abstract

AIMS Scanning laser polarimetry is a new technique allowing quantitative analysis of the retinal nerve fibre layer in vivo. This technique was employed to investigate the variation of the retinal nerve fibre layer thickness in a group of normal subjects of different ages and ethnic groups.

METHODS 150 normal volunteers of different ages and ethnic groups were recruited for this study. Three consecutive 15 degree polarimetric maps were acquired for each subjects. Nerve fibre layer thickness measurements were obtained at 1.5 disc diameters from the optic nerve. Four 90 degree quadrants were identified.

RESULTS The mean nerve fibre layer thickness varied from a minimum of 55.4 μm to a maximum of 105.3 μm, with a mean thickness value of 78.2 (SD 10.6) μm. Superior and inferior quadrants showed a comparatively thicker nerve fibre layer than nasal and temporal quadrants. Retinal nerve fibre layer thickness is inversely correlated with age (p < 0.001). White people showed thicker nerve fibre layers than Afro-Caribbeans (p = 0.002).

CONCLUSION The results indicate a progressive reduction of the nerve fibre layer thickness with increasing age. This may be due to a progressive loss of ganglion axons with age as suggested in postmortem studies. A racial difference in nerve fibre layer thickness is present between whites and Afro-Caribbeans.

  • scanning laser polarimetry
  • retinal nerve fibre layer
  • aging
  • ethnic group

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Aging is accompanied by a reduction of the retinal sensitivity and a deterioration of the visual function.1 2 Some of these changes may be related to a loss of optic nerve fibres with age. Postmortem studies, investigating the optic nerve axon population and its variation with age, have provided considerable evidence of a reduction of the optic nerve axons number with increasing age.3-10

As optic nerve axons and the retinal nerve fibre layer (RNFL) are primarily affected in glaucoma,11 12 age related variations of these structures may be relevant to the assessment of the psychophysical tests commonly used in the diagnosis and follow up of patients with glaucoma, as well as to the morphometrical measurements of the optic nerve head.

In vivo investigation of the RNFL thickness may provide further evidence of ganglion cell axons loss with age.

Although peripapillary RNFL can be directly visualised and imaged in vivo, using red-free fundus ophthalmoscopy and photography,13-15 these techniques allow, so far, subjective, qualitative, and semiquantitative evaluation16 17 of the peripapillary RNFL, but not quantitative thickness measurements.

With the introduction of scanning laser polarimetry, objective measurements of the RNFL thickness may be indirectly estimated in vivo, on the basis of the amount of polarisation change induced by the RNFL on a polarised laser beam passing through it.18 19 This technique may allow us to obtain pointwise measurements, in vivo, of the RNFL thickness over a 15 degree retinal area.20

The aim of this study was to evaluate whether variations of the RNFL thickness were detectable in a population of normal subjects of different ages and ethnic groups using a scanning laser polarimeter (Nerve Fibre Analyzer, Laser Diagnostic Technologies Inc, San Diego, CA, USA).

Materials and methods

Basic principles and technical characteristics of the nerve fibre analyser (NFA) are extensively described elsewhere.18-20The NFA consists of a laser source, a polariser, a scanning unit, a polarisation modulator, a corneal compensator, and a polarisation detector. In principle, a polarised, near infrared laser light (780 nm), travels across the ocular media to the retina and double passes the RNFL before emerging from the eye. The RNFL, behaving as a birefringent medium, changes the state of polarisation of the light passing through it. The polarisation detection unit quantifies the amount of change in the polarisation (retardation) of the light reflected back from the retina, and correlates the amount of retardation with the RNFL thickness. Approximately 1 degree of retardation corresponds to 7.4 μm of RNFL thickness measured histopathologically after removal of the cornea.19Travelling to and from the retina polarised light double passes the cornea. A ‘corneal compensator’ was introduced to correct for the effect of corneal birefringence on the RNFL retardation measurements.20

A total of 65 536 retinal locations are singly tested, allowing us to create a retardation map corresponding to RNFL thickness over a 15 degree (256 × 256 pixels) retinal area with a spatial resolution of approximately 17 × 17 μm.21

For this study normal volunteers of different ages and ethnic groups were recruited from Moorfields Eye Hospital staff and their family, patients’ spouses, and friends.

Inclusion criteria were absence of known ophthalmic pathologies affecting the optic nerve or retina, a negative family history for glaucoma, an intraocular pressure ≤ 21 mm Hg at the time of examination, and a normal ophthalmoscopic appearance of the optic nerve head at biomicroscopy. Subjects with a refractive error exceeding plus or minus 5 dioptres were excluded from the study.

Informed consent was obtained before imaging with the NFA. For each subject only the right eye was selected for the investigation. In order to improve the accuracy and to test the NFA reproducibility, three images were acquired in sequence and subsequently stored in the computer before the analysis. None of the subjects had pupil dilatation before image acquisition.

Images were acquired by one of the authors (DP), with a 1 year experience of using the NFA. A 15 degree field size was used and the optic nerve head was centred in the middle of the monitor LCD display. Light intensity was adjusted for each subject according to the signal level present on the display.

Image analysis was carried out, using NFA software version 2.11.1, and the following standard settings. RNFL thickness was measured over a circular band of 10 pixels width and 1.5 disc diameters. The band was centred around the optic nerve head. Four 90 degrees quadrants divided the measurements in four sectors: superior 45–135 degrees, inferior 225–315 degrees, nasal 135–225, and temporal 45–315 degrees (Fig1).

Figure 1

Fifteen degree retardation map of the retinal nerve fibre layer thickness of the right eye of a 35 year old subject. In the grey scale image thicker parts of retinal nerve fibre layer are indicated with lighter shades and thinner parts with darker shades. A circle band of 10 pixels width is centred around the optic disc at 1.5 disc diameter distance. The topographic map is divided in four 90 degree quadrants. Retinal nerve fibre layer thickness is measured within the circle band.

RNFL thickness values were obtained from the separate analysis of the three images acquired for each subject. NFA reproducibility was evaluated by computing the mean, standard deviation, and coefficient of variation (COV = SD/mean × 100) of the RNFL thickness values obtained for each subject. Analysis of variance (ANOVA) was employed to test the null hypothesis of no difference for the variation of the measurements between subjects of different ages. The NFA measurements and age data were regressed as continuous variables. Ethnic groups were compared using the Mann–Whitney U test. For all the tests a p value less than 0.05 was considered significant.

Results

A total of 150 volunteers of four ethnic groups were recruited for this study. Of these 59 were white (39.3%), 51 Afro-Caribbean (34%), 24 Asian (16%), and 16 Chinese (10.7%). Ages ranged between 5 to 90 years with a mean (1 SD) age of 43.9 (20.9) years. Male/female ratio was 1.3 (85/65). Patients were subdivided in eight groups according to age (Table 1).

Table 1

Subject distribution according to age

The NFA software automatically calculated the average RNFL thickness measured within the 10 pixels width circular band. Thickness values of the four 90 degree quadrants—superior, inferior, nasal, and temporal—were recorded separately. Mean RNFL thickness values for the three images acquired were computed for each subject.

The mean RNFL thickness varied from a minimum of 55.4 μm to a maximum of 105.3 μm, with a mean thickness of 78.2 (10.6) μm. Similar values of RNFL thickness were found between men and women, 78.4 (11.6) and 77.8 (9.2) μm, respectively (p = 0.91). Mean RNFL thickness values for age groups are reported in Table 2.

Table 2

Mean retinal nerve fibre layer (RNFL) thickness (μm) by age groups

Reproducibility was assessed by computing the COV of the three average RNFL thickness values obtained for each subject. The COV varied from a minimum of 0.1% to a maximum of 19.7% with a median value of 3.9% (Table 2). The COV did not vary significantly among the age groups (p = 0.59).

When the mean RNFL thickness and the age were plotted for each subject, the RNFL thickness was seen to decrease significantly (p < 0.001) with increasing age (Fig 2). This applied to all the quadrants, but more noticeably to the inferior and superior quadrants (Fig 3). The topographical distribution of the RNFL thickness among the four quadrants investigated was similar in all age groups. Superior and the inferior quadrants showed highest RNFL thickness values; temporal, and nasal quadrants revealed comparatively thinner values (Fig 3). In order to assess whether the RNFL thickness presented similar variations with age in subjects of different ethnic groups, linear regression analysis was carried out for the white and the Afro-Caribbean groups. Asian and Chinese groups were excluded as the number of subjects in these groups was considered to be too small for the analysis. Both white and Afro-Caribbean groups showed a significant reduction in mean RNFL thickness with increasing age (p < 0.001 for both groups) (Fig 4). Although the rate of decay of RNFL thickness with age was similar in the two groups, in this study Afro-Caribbeans generally showed lower values of RNFL thickness compared with whites. Mean RNFL thickness and age groups for both whites and Afro-Caribbeans is represented in Table3. The RNFL thickness in whites ranged from 60.3 to 105.3 μm with a mean of 79.7 (10.7) μm. The RNFL thickness in Afro-Caribbeans ranged from 71.8 to 94.9 μm with a mean of 79.6 (7.9) μm. The difference between the two ethnic groups was statistically significant (p = 0.002).

Figure 2

Linear regression analysis of the retinal nerve fibre layer thickness and age. R = 0.5938, R2 = 0.3526, SE 8.59324, p = <0.001, n = 150. The analysis indicates a retinal nerve fibre layer thickness decay of 0.38 μm/year.

Figure 3

Mean retinal nerve fibre layer thickness by quadrants for the different age groups.

Figure 4

Linear regression analysis of the retinal nerve fibre layer thickness and age for white and Afro-Caribbean groups. R = 0.6499, R2 = 0.4224, SE 8.20026, p < 0.001, n = 59 for whites and R = 0.7588, R2 = 0.5758, SE 6.37953, p < 0.001, n = 51 for Afro-Caribbeans.

Table 3

Mean retinal nerve fibre layer (RNFL) thickness (μm)by age groups in white and Afro-Caribbean groups

Discussion

Investigations on the effect of age on the optic nerve fibre population in normal individuals has been the subject of various histological studies.3-10 Dolman and colleagues3 described a diminishing density of axons in the optic nerve sections of elderly subjects, accompanied by an increasing presence of fibrous tissue. Balazi and colleagues,4 quantitatively studied the effect of aging on the optic nerve fibre population, reporting a mean loss of 5.637 fibres per year, corresponding to a total loss of about 35% of the optic nerve fibre population in a lifespan. Subsequent studies,5-10 using automated techniques of analysis of the specimens, more accurately determined the number of optic nerve fibres in normal subjects and its variation with age. Besides the differences in mean nerve fibre counts in these studies, a decline of the optic nerve fibre population with increasing age was found with a rate of loss of 4000–5000 fibres per year.4 6 8 10

As the optic nerve axon numbers decrease with age, variations of the RNFL thickness may also be expected.

Evaluation of the RNFL in vivo has been so far limited to qualitative or semiquantitative assessments of red-free fundus photographs16 17 or, more recently, to images of the RNFL acquired with the scanning laser ophthalmoscope.22 Jonaset al 23 correlated the visibility of RNFL, semiquantitatively assessed on red-free photographs, with the age of a large number of normal subjects, describing a reducing visibility of the RNFL with increasing age. As the visibility of the RNFL depends on its thickness, the optical variation of this structure with age may be related to loss of the retinal nerve fibres.

The scanning laser polarimeter was developed to measure the RNFL thickness in vivo by adopting its birefringent property of changing the state of polarisation of a laser light passing through it. In principle, the amount of change in polarisation (retardation) is thought to be proportional to the thickness of the RNFL at any tested point.18 A histopathological study19validating retardation measurements of the RNFL, acquired using the NFA, suggested a good correlation between histological and retardation measurements.

Using the scanning laser polarimeter, we intended to investigate quantitatively in vivo the RNFL thickness in a population of normal volunteers. The results indicate a considerable variation of RNFL thickness among the subjects studied, even of the same age groups. A significant reduction of the RNFL thickness was found with increasing age, with a decay of 0.38 μm/year. Furthermore, retardation measured at different quadrants showed variations corresponding to some of the known topographical characteristics of RNFL described in primate eyes.24 25 Superior and inferior quadrants presented the highest values of RNFL thickness, whereas temporal and nasal quadrants showed consistently lower thickness values.

Similar findings were also described in a previous studies using the scanning laser polarimeter to investigate the RNFL of normal subjects and glaucomatous patients.21 Using the same instrument, Chi et al 26 calculated that the RNFL thickness decreases with a rate of 0.2 μm/year.

The RNFL is not the only birefringent structure of the eyes, scanning laser polarimetric measurements are influenced by the anterior segment birefringence (cornea and lens). Although this source of artefacts should be appropriately compensated in the NFA,20 we cannot exclude that possible variation of the anterior optical media birefringence with age may have somehow influenced our findings.

Both white and Afro-Caribbean groups showed a similar reduction of RNFL thickness with age. However, for the same age groups, RNFL in whites was often found to be thicker than in Afro-Caribbeans. This ethnic difference may be explained by a systematic variation in site of measurements between whites and Afro-Caribbeans as a result of a difference in optic disc size between the two ethnic groups.28 We used the NFA to measure the RNFL thickness at 1.5 disc diameters. This could result in measurements being made further away from the optic nerve head in eyes with large optic discs, with a resulting difference in retardation measurements, as the RNFL thickness diminishes from the optic disc towards the periphery.26 As in this study no optic disc measurements were carried out, we are unable to exclude the possibility that ethnic differences in RNFL thickness may be due to difference in disc size.

However, our findings are supported by those of Varma et al 27 who noted in a study of urban white and black Americans that for optic discs of the same size blacks had a smaller neuroretinal rim area than whites.

In conclusion, our study further supports the evidence of a progressive decay of ganglion fibres with age. Furthermore, a difference in RNFL thickness between white and Afro-Caribbeans, measured by scanning laser polarimetry, may indicate ethnic differences in nerve fibres in the normal population.