Article Text

Download PDFPDF

Quantitative evaluation of fundus autofluorescence imaged “in vivo” in eyes with retinal disease
  1. Noemi Loisa,b,c,
  2. Anthony S Halfyarda,
  3. Alan C Birda,b,
  4. Frederick W Fitzkea
  1. aDepartment of Visual Science, Institute of Ophthalmology, London, bMedical Retina Service, Moorfields Eye Hospital, London, cVitreo-Retinal Service, St Paul's Eye Unit, The Royal Liverpool University Hospital, Liverpool
  1. Professor Frederick W Fitzke, Institute of Ophthalmology, Londonf.fitzke{at}ucl.ac.uk

Abstract

AIM To describe a new method of evaluating the topographic distribution of fundus autofluorescence in eyes with retinal disease.

METHODS Images of fundus autofluorescence were obtained in five patients and 34 normal volunteers using a confocal scanning laser ophthalmoscope (cSLO). To evaluate the topographic distribution of fundus autofluorescence throughout the posterior pole a rectangular box, 10 × 750 pixels, was used as the area of analysis. The box was placed, horizontally, across the macular region. The intensity of fundus autofluorescence of each pixel within the rectangular box was plotted against its degree of eccentricity. Profiles of fundus autofluorescence from patients were compared with those obtained from the age matched control group and with cSLO images.

RESULTS Profiles of fundus autofluorescence appeared to represent the topographic distribution of fundus autofluorescence throughout the posterior pole appreciated in the cSLO images, and allowed rapid identification and quantification of areas of increased or decreased fundus autofluorescence.

CONCLUSIONS Fundus autofluorescence profiles appear to be useful to study the spatial distribution of fundus autofluorescence in eyes with retinal disease.

  • autofluorescence
  • lipofuscin
  • retinal imaging
  • scanning laser ophthalmoscope

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

The study of accumulation of lipofuscin in the retinal pigment epithelium (RPE) has attracted the attention of many researchers in the past decades. Although still incompletely understood, it is believed that lipofuscin represents the end product of phagocytosed photoreceptor outer segment discs, and to a lesser extent, of autophagy.1-6 “Ex vivo” studies in donor human eyes have shown that lipofuscin accumulates in the RPE with age7-9 and retinal disease.10-18 Similarly, it has been demonstrated that, in normal subjects, there is a topographic distribution of RPE lipofuscin content throughout the fundus, being higher at the posterior pole, with a localised dip at the fovea, and lower towards the peripheral retina.7-9

Excessive accumulation of lipofuscin in the RPE has been associated with degeneration of RPE cells and photoreceptors.17 19Hence, information regarding the distribution of fundus autofluorescence may provide insight on the sequence of events that lead to retina damage in a variety of retinal diseases. Until recently, information regarding the content and spatial distribution of RPE lipofuscin in eyes with retinal disease was limited by the scarce number of eyes available for histopathological evaluation. This problem has been recently overcome with the development of new techniques that allow the study of RPE lipofuscin levels “in vivo”.20-25 Using a spectrophotometer20-22 and a confocal scanning laser ophthalmoscope (cSLO)23-25 the level of fundus autofluorescence (index of RPE lipofuscin content) has been assessed “in vivo” in a variety of retinal disorders. Previous in vivo studies have quantified fundus autofluorescence in eyes with retinal disease at selected areas of the retina.20-25 These isolated measurements have provided valuable information in a number of retinal disorders. However, the differential distribution of fundus autofluorescence, appreciated in the images obtained using a cSLO,23-25 can not be represented by an isolated measurement.

The aim of the present study was to design a method of representing the topographic distribution of fundus autofluorescence in eyes with retinal disease. In order to test this technique, fundus autofluorescent studies were performed in a group of patients with inherited retinal disorders and in an age matched control group.

Subjects and methods

Five patients with retinal disease and 34 normal volunteers were included in the study. The median age of patients was 29 years (range 20–36 years), and of normal volunteers 30 years (range 20–36 years). One patient with each of Best disease, adult vitelliform foveomacular dystrophy, Stargardt macular dystrophy-fundus flavimaculatus, bull's eye macular dystrophy, and cone-rod dystrophy were included in the study. Fundus examination of normal volunteers disclosed no abnormalities. All subjects had clear lenses on slit lamp biomicroscopy.

In all subjects one eye was selected and levels of fundus autofluorescence were recorded. Images of fundus autofluorescence were obtained, after pupillary dilatation, using a confocal scanning laser ophthalmoscope (cSLO, Zeiss, Oberkochen, Germany), as has been previously described.26 Briefly, autofluorescent images were produced using a 40 degree field of view mode, and a confocal aperture which provided a full width half maximum depth resolution of less than 600 μm. The ametropic corrector was employed to correct for refractive error and to focus on the structure of interest. An argon laser light (488 nm) was used for illumination, and a wide band pass filter with short wavelength cut off (521 nm) was inserted in front of the detector to record fundus autofluorescence. The argon laser was measured at regular intervals and found to have a constant power output of 250 μW at the cornea. The automatic gain control was switched off and the instrument gain was kept constant for all measurements of normal volunteers and patients. A series of images of the eye studied was recorded at standard video scanning rates on VHS video tape. Later, 32 consecutive frames were digitised, aligned, and averaged to obtain each image. Since all normal volunteers and patients had clear lenses, no correction for lens opacities was made

In order to analyse the topographic distribution of fundus autofluorescence the following computer program was designed and written. A rectangular box, 10 pixels in height by 750 pixels in width, was used as the area of analysis. The box was placed, horizontally, across the macular region with its centre located at the centre of the fovea. In patients in whom the centre of the fovea was difficult to recognise its location was estimated based on its expected anatomical location. The centre of the fovea was considered as 0 degrees of eccentricity. The length of the box allowed the study of fundus autofluorescence in an area extending from the optic nerve to the temporal aspect of the macula (12–16 degrees of eccentricity). For each subject the coordinates corresponding to the centre of the fovea, the horizontal line corresponding to the area of analysis, and the nasal and temporal ends of the area studied were marked and saved in the fundus autofluorescent image (Fig 1).

Figure 1

(Left top) Fundus photograph (red-free) of a 20 year old woman with Best's disease. (Right top) Fundus autofluorescent image of the same patient. The intersection between the horizontal and vertical lines represents the centre of the fovea. The horizontal line shows the area where measurements were taken. The right and left ends of the area of analysis are also marked on the image. (Right bottom) Fundus autofluorescent profile (dotted graph). Dark dots represent values of fundus autofluorescence of single pixels within the area studied. The highest, mean, and lowest values of fundus autofluorescence are represented by continuous dark lines (top, medium, and bottom lines, respectively). A graph with values of fundus autofluorescence corresponding to an age matched control group is also shown. Two light grey bands represent the 25% (inferior grey band) and 75% (superior grey band) confident interval. The mean (black line), lowest (inferior aspect of the dark grey band), and highest (superior aspect of the dark grey band) values of fundus autofluorescence at each degree of eccentricity are also represented. The regional increased autofluorescence is localised to the centre of the macula.

The intensity of fundus autofluorescence was measured in grey scale arbitrary units. The value of fundus autofluorescence of each pixel within the rectangular box was plotted against its degree of eccentricity with respect to the fovea, after having subtracted the video offset grey scale measured over the optic nerve head. Thus, the autofluorescence intensity of each of the 10 pixels within the width of the box at a particular degree of eccentricity was represented in a graph as an individual point depending on its value of autofluorescence. In patients, the highest, mean, and lowest values of each of the 10 pixels throughout the entire area studied were indicated in the graph as continuous lines (Fig 1).

A plot was generated with the 25% and 75% confidence interval (lower and upper quartiles) and the mean of the mean, lowest and highest values obtained at each degree of eccentricity for all normal volunteers. Single plots of fundus autofluorescence obtained from each patient were compared with cSLO images and with the plot obtained from the age matched control group (Figs1, 2, 3, 4, 5).

Results

In patients with retinal disease, topographic plots demonstrated well the spatial distribution of fundus autofluorescence throughout the posterior pole seen in cSLO images (Figs 1, 2, 3, 4, 5). Thus, high levels of fundus autofluorescence detected at the centre of the macula in a patient with Best's disease corresponded well with the intense signal from the “pseudohypopyon” lesion observed in the cSLO image (Fig1). Similarly, high levels of fundus autofluorescence were observed at the centre of the fovea in a patient with adult vitelliform foveomacular dystrophy corresponding with an area of intense signal in the cSLO image (Fig 2). In a patient with Stargardt macular dystrophy-fundus flavimaculatus several peaks of increased fundus autofluorescence appeared to correspond well with foci of increased signal from “flecks” observed in the cSLO image (Fig 3). In a patient with bull's eye macular dystrophy no focal changes in fundus autofluorescence were detected outside the centre of the macula (Fig4). In a patient with cone-rod dystrophy low levels of fundus autofluorescence were measured at the macula corresponding to a large area with low intensity signal in the cSLO image (Fig 5).

Figure 2

(Left top) Fundus photograph (red-free) of a 36 year old man with adult vitelliform foveomacular dystrophy. Fundus autofluorescent image (right top) and fundus autofluorescent profile of the same patient (right bottom). Very high levels of fundus autofluorescence were detected at the centre of the fovea. Background values of fundus autofluorescence were also high throughout the entire area studied.

Figure 3

Fundus autofluorescent image of a 35 year old woman with Stargardt macular dystrophy-fundus flavimaculatus (top). An area of decreased signal at the centre of the macula and small foci of increased signal within it and around it were seen. Fundus autofluorescence profile showed very high background fundus autofluorescence levels across the entire area studied. Values of fundus autofluorescence at the fovea were within the lower normal values and peaks of increased autofluorescence at that site were also detected (bottom).

Figure 4

Fundus autofluorescent image of a 28 year old female with a bull's eye macular dystrophy (top). No focal areas with high or low intensity signal were detected outside the fovea in the cSLO image. Fundus autofluorescence profile demonstrated very high levels of fundus autofluorescence throughout the posterior pole (bottom).

Figure 5

Fundus autofluorescence profile in a 32 years old man with cone-rod dystrophy. A large area with low intensity signal was observed at the macula in the cSLO image (top). This area appeared to be surrounded by a ring of high intensity signal. The fundus autofluorescence profile disclosed levels of fundus autofluorescence at the centre of the macula within the lower normal range and high levels of fundus autofluorescence outside the centre.

Topographic plots allowed also the study of background fundus autofluorescence, measured outside areas of focal increased or decreased fundus autofluorescence. Background values of fundus autofluorescence were found to be high throughout the posterior pole in patients with adult vitelliform foveomacular dystrophy, Stargardt macular dystrophy -fundus flavimaculatus, and bull's eye macular dystrophy (Figs 2, 3, 4).

Discussion

There appears to be a clear value in studying the distribution of fundus autofluorescence in eyes with retinal disorders. Hence, it may provide a greater understanding of the pathogenetic mechanisms of retinal disease. Similarly, it may explain the preferential regional involvement observed in certain retinal disorders, such as the perifoveal involvement that occurs in early stages of age related macular degeneration. In addition, the distribution of fundus autofluorescence may be helpful to predict the extension of further retinal damage—that is, high levels of fundus autofluorescence involving the posterior pole may indicate the development of a large area of atrophy whereas a small restricted area of increased autofluorescence may suggest a more favourable outcome.

In previous reports, Delori and associates measured fundus autofluorescence at 7 degrees temporal to the fovea and at the fovea in eyes with retinal disease, where the highest and lowest levels of fundus autofluorescence have been found in normal volunteers.20-22 Von Ruckmann and colleagues used as the area of analysis an 8 × 8 pixel box located anywhere in the fundus where no focal changes were observed to obtain background levels of autofluorescence.24 25 Although studying levels of fundus autofluorescence at 7 degrees temporal to the fovea and at the fovea provides valuable information and allows comparison between individuals, it remains unclear if those locations are the most appropriate to evaluate fundus autofluorescence in all retinal diseases. Similarly, measuring fundus autofluorescence in a selected area of the fundus has several limitations. Firstly, it may not allow accurate comparisons between patients, since the eccentricity selected to measure fundus autofluorescence may be different in different individuals. Secondly, if it is proposed to study changes in fundus autofluorescence over time, autofluorescence measurements should be obtained always at the same area for all follow up visits, which may be difficult if this approach is used. Lastly, the differential distribution of fundus autofluorescence throughout the fundus, appreciated in the images obtained using a cSLO,23-25 can not be represented by an isolated measurement.

In this study, topographic plots were used to represent the differential distribution of fundus autofluorescence in eyes with retinal disease. This method of analysis allowed measurement of fundus autofluorescence throughout a large area of the posterior pole and provided a good estimate of variations in fundus autofluorescence across this region. Quantitative values of fundus autofluorescence could be compared with the qualitative findings observed in the cSLO images. Similarly, fundus autofluorescent profiles obtained from patients could be compared with those obtained from an age matched control group, allowing an estimation of areas of increased/decreased fundus autofluorescence. In addition, since the software used records the coordinates corresponding to the area of analysis, it would be possible to evaluate changes of fundus autofluorescence in the area studied over time.

The purpose of plotting values of fundus autofluorescence of single pixels throughout the entire area studied instead of the average value of a certain number of pixels, was to avoid the smoothing effect in the values of autofluorescence that would occur if the latter approach was used. Plotting values of autofluorescence of single pixels allows a better representation of small focal changes in fundus autofluorescence that may occur in some retinal diseases, such as those that be observed in patients with Stargardt macular dystrophy-fundus flavimaculatus.

Although cSLO autofluorescent images provide qualitative information on the topographic changes of fundus autofluorescence, quantitative data appear to be important in several circumstances. Thus, quantitative data are needed when levels of background fundus autofluorescence are to be evaluated. Similarly, it may be useful in determining if a particular area has a low level of fundus autofluorescence with respect to background or if levels of autofluorescence in that area are normal within the context of high background fundus autofluorescence.

Acknowledgments

Supported by the Macular Disease Society, United Kingdom (NL), and the Foundation Fighting Blindness, United States.

References