Article Text

Download PDFPDF
Trabecular meshwork bypass tube shunt: initial case series
  1. D Spiegel,
  2. K Kobuch
  1. Department of Ophthalmology, University of Regensburg, Regensburg, Germany
  1. Correspondence to: Professor Detlev Spiegel, Klinik und Poliklinik für Augenheilkunde der Universität Regensburg, Franz-Josef-Stras-Allee 11, 93042 Regensburg, Germany; detlev.spiegel{at}klinik.uni-regensburg.de

Abstract

Aims: This study describes a prospective consecutive case series of the initial six eyes of five patients undergoing implantation of the trabecular bypass tube shunt.

Methods: A prospective consecutive case series. The initial six eyes of five patients with uncontrolled open angle glaucoma who had never previously undergone ocular surgery. Implantation of the trabecular bypass tube shunt measuring 150 μm outer diameter and 50 μm inner diameter was performed with the distal end placed in Schlemm’s canal and the proximal end in the anterior chamber. The main outcome measures were visual acuity, intraocular pressure, glaucoma medication use.

Results: The tube was successfully implanted in five of six eyes. In four eyes longer term follow up of 5–9 months showed no loss of visual acuity with decreased intraocular pressure from preoperative levels (mean 23.4–16.5 mm Hg) and reduced requirement of glaucoma medications (mean 3–0.5). In a subset of two eyes, there was no measured increase in aqueous flare or reduction of endothelial cell count. In one eye the tube was explanted because of presumed misplacement by excessive bleeding during surgery. Two eyes showed a diffuse bleb.

Conclusions: This study reports the initial experience with a novel approach to surgical glaucoma therapy. This very small tube allows a direct communication to be established between the anterior chamber and Schlemm’s canal, effecting a trabecular bypass. In this small number of eyes this procedure reduced intraocular pressure and the need for glaucoma medications without appreciable side effects.

  • trabecular meshwork bypass tube shunt

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • Proprietary or commercial interests: None