Non-invasive volumetric blood flow measurement in the orbit
  1. J Németh1,
  2. Z Harkányi2,
  3. F Humml2
  1. 11st Department of Ophthalmology, Semmelweis University, Budapest, Hungary
  2. 2Department of Radiology, Heim Pal Children Hospital, Budapest, Hungary
  1. Correspondence to: Dr János Nemeth, 1st Department of Ophthalmology, Semmelweis University, Tömõ u 25–29, H-1083, Budapest, Hungary; nj{at}szem1.sote.hu

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

We read with great interest the article of Orge and co-workers1 who claim a first ever solution to the problem of non-invasive volumetric blood flow measurement in the ophthalmic artery. This is a very important topic both from the clinical and scientific point of view, since blood supply is an important factor, for example, in glaucoma studies. However, to date for the orbital circulation only the blood velocity has been measurable; from this the resistivity indices can be calculated, but the volumetric flow cannot be determined because the small diameter of the orbital vessels does not allow the vessel diameter measurement, which is indispensable for such calculation.

In spite of the novelty of the report and the extensive analysis of the authors on the possible sources of high variability of volumetric flow measurements, it appears to us that there are several points which need further consideration; some of these were analysed in the editorial by Hedges,2 but we would like to highlight two …