Article Text
Abstract
Aims: To devise and test strategies for improving Heidelberg retina tomograph (HRT and HRT-II) rim area (RA) repeatability and assess the benefit of the strategies in time series of HRT images.
Methods: The effect of the standard and 320 μm reference planes and image quality on RA repeatability was assessed in a test-retest HRT image dataset from 74 subjects. A longitudinal HRT image dataset from 30 ocular hypertensive subjects was analysed by linear regression of RA over time, with each of the reference planes and using a manual image alignment facility. RA variability was estimated by comparing the standard deviation of residuals (RSD) generated by each linear regression.
Results: RA repeatability was better with the 320 μm reference plane (repeatability coefficient 0.17 mm2), improving further with only good quality images (repeatability coefficient 0.08 mm2). For the longitudinal data, a significant (p<0.0001) reduction in the RSD from 0.10 to 0.05 mm2 was obtained with the 320 μm reference plane. Manual alignment led to a further significant (p<0.0001) reduction in the RSD to 0.04 mm2.
Conclusions: The findings support the use of a 320 μm reference plane and manual image alignment to analyse RA over time. The estimates of RA repeatability may be used to define thresholds for glaucomatous change.
- HRT, Heidelberg retina tomograph
- OHT, ocular hypertension
- ONH, optic nerve head
- POAG, primary open angle glaucoma
- RA, rim area
- RSD, residual standard deviation
- VF, visual field
- Heidelberg retina tomograph
- glaucoma
- HRT, Heidelberg retina tomograph
- OHT, ocular hypertension
- ONH, optic nerve head
- POAG, primary open angle glaucoma
- RA, rim area
- RSD, residual standard deviation
- VF, visual field
- Heidelberg retina tomograph
- glaucoma