Comparison of optical coherence tomography models OCT1 and Stratus OCT for macular retinal thickness measurement
Free
  1. V Pierre-Kahn,
  2. R Tadayoni,
  3. B Haouchine,
  4. P Massin,
  5. A Gaudric
  1. Department of Ophthalmology, Hospital Lariboisiere, Assistance Publique-Hôpitaux de Paris, Université Paris 7, Paris, France
  1. Correspondence to: Ramin Tadayoni MD, Service d’Ophtalmologie, Hôpital Lariboisière 2, rue Ambroise Paré, 75475 Paris Cedex 10, France; ramin.tadayonilrb.ap-hop-paris.fr

Abstract

Aims: To compare the values measured for retinal macular thickness with the first and last generations of the optical coherence tomograph (OCT1 and Stratus OCT, Zeiss, Humphrey Division).

Methods: This was a cohort study. 59 eyes were examined: 17 had a normal macula and 42 had a diabetic macular oedema. In each eye, mean retinal thickness (RT) was measured automatically in the nine macular Early Treatment Diabetic Retinopathy Study areas and at the foveal centre, using OCT1 and Stratus OCT. The paired mean RT values for each area and the type and proportion of artefacts were compared.

Results: Of the 590 automatic measurements, 505 had no artefact, either with OCT1 or Stratus OCT. The mean difference between the OCT1 and Stratus OCT measurements was 25 (SD 26.2) μm (p<0.0001). With Stratus OCT, RT values were significantly higher, by 8.1% (7.8%), than with OCT1. Artefacts were only observed in cases of diabetic macular oedema and were significantly more frequent with OCT1 than Stratus OCT (10.5% versus 4.4, p<0.0001).

Conclusion: The macular retinal thickness values measured with Stratus OCT were significantly higher than those measured with OCT1. Stratus OCT has the advantage of producing fewer artefacts than OCT1 in pathological cases.

  • IS, inner segment
  • OCT, optical coherence tomography
  • OS, outer segment
  • RPE, retinal pigment epithelium
  • RT, retinal thickness
  • retina
  • optical coherence tomography
  • macular oedema
  • retinal thickness
  • IS, inner segment
  • OCT, optical coherence tomography
  • OS, outer segment
  • RPE, retinal pigment epithelium
  • RT, retinal thickness
  • retina
  • optical coherence tomography
  • macular oedema
  • retinal thickness

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes