The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry
  1. P-A Tonnu1,
  2. T Ho1,
  3. T Newson2,
  4. A El Sheikh2,
  5. K Sharma1,
  6. E White1,
  7. C Bunce3,
  8. D Garway-Heath1
  1. 1Glaucoma Research Unit, Moorfields Eye Hospital, London EC1V 2PD, UK
  2. 2Division of Civil Engineering, Faculty of Engineering, University of Dundee, Dundee DD1 4HN, UK
  3. 3Department of Research and Development, Moorfields Eye Hospital, London EC1V 2PD, UK
  1. Correspondence to: D F Garway-Heath MD, FRCOphth, Glaucoma Research Unit, Moorfields Eye Hospital, London EC1V 2PD, UK; david.garway-heathmoorfields.nhs.uk

Abstract

Aims: To evaluate the influence of central corneal thickness (CCT) on intraocular pressure (IOP) measurements made with the Goldmann applanation tonometer (GAT), Tono-Pen XL, ocular blood flow tonograph (OBF), and Canon TX-10 non-contact tonometer (NCT).

Methods: CCT was recorded for either eye (randomly selected) of each of 105 untreated patients with ocular hypertension and glaucoma attending the glaucoma research unit at Moorfields Eye Hospital. For each of the selected eyes, IOP was measured with the GAT (two observers), Tono-Pen, OBF, and NCT in a randomised order. The relation of measured IOP and of inter-tonometer differences with CCT and subject age was explored by linear regression analysis.

Results: A significant association between measured IOP and CCT was found with each instrument. The change in measured IOP for a 10 μm increase in CCT was 0.28, 0.31, 0.38, and 0.46 for the GAT, Tono-Pen, OBF, and NCT, respectively (all p⩽0.05). There was a significant association between the NCT/GAT differences and CCT, with a tendency of NCT to overestimate GAT in eyes with thicker corneas. There was a significant association between GAT/Tono-Pen and OBF/Tono-Pen differences and age, with a tendency of GAT and OBF to overestimate the Tono-Pen in eyes of older subjects.

Conclusion: IOP measurement by all four methods is affected by CCT. The NCT is affected by CCT significantly more than the GAT. Subject age has a differential effect on the IOP measurements made by the GAT and OBF compared to the Tono-Pen.

  • CCT, central corneal thickness
  • DCT, dynamic contour tonometer
  • GAT, Goldmann applanation tonometer
  • IOP, intraocular pressure
  • NCT, non-contact tonometer
  • OBF, ocular blood flow
  • ORA, ocular response analyser
  • tonometry
  • intraocular pressure
  • cornea
  • age
  • measurement error
  • CCT, central corneal thickness
  • DCT, dynamic contour tonometer
  • GAT, Goldmann applanation tonometer
  • IOP, intraocular pressure
  • NCT, non-contact tonometer
  • OBF, ocular blood flow
  • ORA, ocular response analyser
  • tonometry
  • intraocular pressure
  • cornea
  • age
  • measurement error

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes