Baseline IOP predicts selective laser trabeculoplasty success at 1 year post-treatment: results from a randomised clinical trial
Free
  1. W G Hodge,
  2. K F Damji,
  3. W Rock,
  4. R Buhrmann,
  5. A M Bovell,
  6. Y Pan
  1. Department of Ophthalmology, The University of Ottawa Eye Institute, 501 Smyth Road, Ottawa, Ontario, K1H 8L6 Canada
  1. Correspondence to: Dr William G Hodge The University of Ottawa Eye Institute, 501 Smyth Road, Ottawa, ON K1H 8L6 Canada; whodgeottawahospital.on.ca

Abstract

Background/aims: The efficacy and safety of selective laser trabeculoplasty (SLT) has been found to be equivalent to argon laser trabeculoplasty (ALT). Since SLT produces significantly less disturbance to the trabecular meshwork and is theoretically more repeatable than ALT, it has potential to replace ALT as the standard procedure to treat medically uncontrolled open angle glaucoma. This study’s objective is to determine factors that predict successful SLT at 1 year post-treatment.

Methods: As part of a randomised clinical trial comparing the efficacy and safety of SLT to ALT, data on 72 SLT patients were collected, and successful SLT defined as having an SLT induced intraocular pressure (IOP) reduction of ⩾20% at 1 year post-treatment follow up.

Results: 43 out of the 72 patients who had completed their 1 year follow up visit had an IOP reduction of ⩾20% from baseline. No glaucoma risk factors studied predicted successful SLT. The amount of trabecular meshwork pigmentation was not a significant predictor. However, it was discovered that baseline IOP strongly predicted SLT success (odds ratio = 1.16; p = 0.0001).

Conclusion: SLT success was significantly predicted by baseline IOP but not by age, sex, other glaucoma risk factors, type of open angle glaucoma, or by degree of trabecular meshwork pigmentation.

  • ALT, argon laser trabeculoplasty
  • BCVA, best corrected visual acuity
  • IOP, intraocular pressure
  • PI, peripheral laser iridotomy
  • SLT, selective laser trabeculoplasty
  • TM, trabecular meshwork
  • selective laser trabeculoplasty
  • intraocular pressure
  • predictors
  • glaucoma risk factors
  • ALT, argon laser trabeculoplasty
  • BCVA, best corrected visual acuity
  • IOP, intraocular pressure
  • PI, peripheral laser iridotomy
  • SLT, selective laser trabeculoplasty
  • TM, trabecular meshwork
  • selective laser trabeculoplasty
  • intraocular pressure
  • predictors
  • glaucoma risk factors

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Argon laser trabeculoplasty (ALT) is a commonly used method to treat medically uncontrolled open angle glaucoma. Wise and Witter first introduced ALT in 19791; it is believed that laser trabeculoplasty causes increased outflow facility by either or both of the following two mechanisms; (i) mechanical: by tightening the trabecular meshwork (TM) to open the adjacent, untreated inter-trabecular spaces,2 and/or (ii) cellular: by inducing cell division and migration of macrophages to clear the TM debris.3,4

ALT treatment has an impressive intraocular pressure (IOP) lowering effect in patients with elevated IOP, with previously reported early IOP reductions up to 35% from baseline1,5–9; however, its effect declines with time.1,8,10–12 Ten year follow up data showed failure rate of 68–95%,10,11 and repeat ALT was generally perceived as ineffective at maintaining normal IOP as the rate of 1 year success (IOP less than 3 mm from baseline) is 21%.13,14 The high long term failure rate of ALT could be explained by the fact that ALT causes detrimental disruption to the microstructure of the TM, and excessive TM damage determines treatment failure. Histopathological studies revealed remarkable uveoscleral meshwork destruction post-ALT treatment with coagulative tissue necrosis and fragmented cells,15 crater formation,16 heat damaged collagen fibres in tissues,16,17 and membranes surrounding the argon laser spots formed by migrating endothelial cells.12,18–21 It has been suspected that these membranes covering the TM would cause outflow reduction later, leading to treatment failure. Moreover, the remarkable disruption of the TM and accumulation of tissue debris in the juxtacanalicular region are thought to cause the acute IOP elevation seen in some patients after ALT treatment.15

Selective laser trabeculoplasty (SLT) is an alternative laser treatment introduced by Latina et al in 1995.22 SLT is a Q switched, frequency doubled Nd:YAG laser (λ = 532 nm) that selectively targets the pigmented TM cells without adversely affecting the TM in vitro, rendering the TM architecture more preserved.22 There have been a number of studies that compared the efficacy of ALT and SLT based on post-treatment IOP reduction, and all reported that SLT is as effective as ALT in terms of IOP lowering.5,7,9 In general, both modalities lower IOP an average of 5 mm of mercury 6 months post-treatment. Furthermore, compared to ALT, SLT did not cause ablation craters at the border of pigmented and non-pigmented cells in the TM,16 and the cellular changes induced by SLT did not extend beyond the Schlemm’s canal17 as it would after ALT. In addition, SLT appears not to cause the membrane formed by migrating endothelial cells in the necrotic TM seen after ALT treatment.19 SLT allows the use of 80–100 times lower levels of energy and less laser spots on the TM, causing less damage to the TM. Based on the above observations and results, SLT appears to be less destructive and may be more repeatable clinically than ALT.

The focus of current SLT studies is SLT’s potential long term IOP reduction effect, predictors of SLT success and adverse events from SLT in the long term. There are already a number of studies devoted to observe immediate or monitor long term post-treatment IOP reduction; all concluded that SLT is able to maintain satisfactory IOP reduction within 6 months,5,7,20,23–26 at 1 year,27–29 or longer28,30,31 post-treatment. For example, 1 year IOP lowering seems to be maintained at around 5 mm Hg from baseline.27 The most frequently used criterion for successful laser induced IOP lowering was a reduction of at least 20% from baseline prelaser IOP. On the other hand, regarding the predictors of SLT success, only one study attempted to determine the predictors of successful SLT.25 SLT’s success rate varied from 40% to 84% in patient groups of different characteristics.24,25,28,29,31 Thus, there could be indicators that predict better SLT outcome. Significantly higher IOP reduction was found in patients with ⩾22 mm Hg baseline IOP in a prospective non-randomised trial,32 but no prognostic importance of this finding has been suggested. Kano, et al25 discovered that lower baseline IOP is a significant success predictor, whereas factors such as, age, sex, previously failed ALT, and goniopigmentation were found to be non-significant. Similarly, Odberg and Sandvik6 indicated significantly higher risk for failure when a patient’s baseline IOP was ⩾32 mm Hg. Results of age and degree of angle pigmentation in relation to SLT success were mixed.20,25,31

The objective of the present study is to determine the factors that predict successful SLT at 1 year post-treatment.

MATERIALS AND METHODS

Details of patient recruitment, materials, and methods are given in a previous publication of this study.5,33 Briefly, the study took place at the University of Ottawa Eye Institute and was approved by the Ottawa Hospital research ethics board. Potential study subjects were drawn from the patient pool of the institute’s glaucoma clinic, from March 1997 to March 2003. Our inclusion criteria required the study patients to: (i) present with any type of open angle glaucoma including pseudoexfoliation and pigmentary glaucoma; (ii) have uncontrolled IOP of ⩾16 mm Hg with maximal medical therapy (three or more antiglaucoma drugs tolerated by the patient) or had failed previous 180/360 degree ALT treatment; (iii) be over 18 years of age; and (iv) have two sighted eyes. Subjects were excluded if s/he (i) had a type of glaucoma other than open angle; (ii) had advanced visual field defect; (iii) had any previous ocular surgery other than ALT or peripheral laser iridotomy (PI) done in the study eye, or required any ocular surgery within 6 months of this study enrolment; (iv) had cornea disease precluding accurate applantation tonometry; or (v) was or would be on systemic steroids during the study period. All patients were required to sign the informed consent at enrolment. Follow ups were scheduled to take place at 1 week, 1, 3, 6, and 12 months after treatment. Intraocular pressure, best corrected visual acuity (BCVA), anterior chamber reaction, TM pigmentation grade, and number of medications used were recorded at all visits.

A total of 176 patients were recruited and randomised in blocks of six using a computer generated random number list to receive either ALT or SLT. In the end, SLT was done in 89 randomised patients, who constituted the sample under study in this paper. Figure 1 gives the outcome of recruitment, randomisation, and follow up at 1 year post-treatment.

Figure 1

 Outcome of randomisation and follow up to 1 year post-treatment.

Statistical analysis

The primary outcome of interest was SLT induced IOP reduction; successful SLT was defined as having a reduction of ⩾20% of pretreatment IOP at 1 year post-treatment. Best corrected visual acuity and IOP change between the “Success” and “Non-success” group were compared throughout the 1 year follow up period. Other potential predictors studied were age, sex, type of glaucoma, glaucoma risk factors, TM pigmentation, total energy used, and previous ALT. Distribution of these variables was compared between the success and non-success groups. For normal continuous variables, appropriate Student’s t tests were used to detect statistical significance; for categorical variables χ2 tests and the non-parametric Fisher’s exact test; and for nominal categorical variables the Kruskal-Wallis test. Any covariates that achieved p<0.10 for the bivariate analyses were included in the final multivariate logistic regression model. Multi-collinearity was checked before determining the final model. All statistical analyses were performed using the Stata 7.0 software.

RESULTS

Out of the 89 patients who received SLT (table 1), 72 completed their 1 year follow up visit; 43 of which had achieved an IOP reduction of ⩾20% from their baseline IOP. Seventeen patients were lost to follow up because of moving, unwillingness to return for further examinations, sickness, or death.

Table 1

 Baseline characteristics of the total study sample

The success group has considerably higher baseline and 1 hour post-treatment IOP than the non-success group. There are no significant differences in the distribution of glaucoma risk factors and other prognostic factors except for hypertension and family history of glaucoma, which were only marginally significant between groups (p<0.1) (table 2). Four out of the 72 patients (6%), two in each outcome group had an IOP spike defined as having ⩾6 mm Hg rise in IOP from baseline at 1 hour post-treatment.

Table 2

 Comparison of the success and non-success groups (SD)

The baseline IOP for the success group ranged from 18–36 mm Hg and for the non-success group 16–28 mm Hg. A clear trend of steady post-treatment IOP reduction was observed in the success group, whereas in the non-success group there was no significant improvement in terms of IOP reduction throughout the 12 months (table 3). In both outcome groups, patients with lower baseline IOP had lower IOP at 1 year post-treatment (success: p = 0.0002; non-success: p = 0.0001). On the other hand, it was observed in the success group that patients with higher baseline IOP had significantly greater IOP reduction induced by SLT (p = 0.00001), but not in the non-success group (p = 0.2292). The best corrected visual acuity was similar in both groups at baseline and throughout the 1 year follow up. As for the effect of previous ALT treatment, there is no significant difference in IOP regardless of the patients’ success status at 1 year, except at 6 months post-treatment (p = 0.0444).

Table 3

 Post-treatment IOP* throughout the 1 year follow up period

Table 4 provides the simple and multivariate logistic models fitted to obtain the odds ratios (ORs) of the predictor variables. In simple logistic models, both baseline IOP and IOP at 1 hour post-treatment appeared to be significant predictors, but the significance of IOP at 1 hour post-treatment diminished in the multivariate model; multi-collinearity was suspected, since baseline IOP and IOP at 1 hour post-treatment are highly correlated (r = 0.7865, p = 0.00001).

Table 4

 Multivariate logistic regression analysis result

The final multivariate logistic model provided an OR of 1.58 for baseline IOP as a predictor of SLT success; the OR could be interpreted as given that other variables in the multivariate model were controlled for, for every 5 mm Hg increase in baseline IOP, the odds of having successful SLT treatment outcome increases 10 times.

DISCUSSION

To the best of our knowledge, our study is the first highly successful SLT predictor study derived from high quality data of a randomised clinical trial. The results suggest that SLT success does not depend on age or sex. None of the glaucoma risk factors were found to be significant predictors. Family history of glaucoma was the closest to being significant, but since the analysis was not adequately powered to detect a difference, we cannot draw conclusions based on these data.

Interestingly, unlike ALT, TM pigmentation and type of glaucoma did not predict better outcome. Our results agree with previous findings that lower baseline IOP predicts lower post-treatment IOP,5,6,25 and patients with higher baseline IOP benefit more from the laser treatment in terms of total IOP reduction.32 Unlike previous studies where arbitrary cut offs of high baseline IOP were used in the analysis, we demonstrated a continuous relation between baseline IOP and post-treatment IOP reduction to be significant. Since our definition of success is based on the patients’ relative response to the SLT treatment, we expect to observe most patients in the success group to have higher baseline IOP. As reported in the results section, the non-success group had a narrower range of baseline IOP than the success group, and all non-success patients had a baseline IOP of <29 mm Hg; thus, the data might seem to indicate a regression towards the mean effect. However, the fact that even after all patients with IOP ⩾30 mm Hg at baseline were excluded in a separate analysis, the results still suggested that higher baseline IOP predicted better chance of SLT outcome in terms of post-treatment IOP reduction.

Our result regarding predictors of successful SLT are comparable to previous studies on successful ALT. Baseline IOP was found to correlate with IOP reduction post-ALT as well.34–37 Lower baseline IOP predicts lower post-ALT IOP, and higher baseline IOP predicts greater IOP reduction post-ALT. Interestingly, in ALT studies it was also found that presence of pseudoexfoliation,34 ALT performed as primary therapy,36,38 grade of pigmentation,38,39 and number of preoperative antiglaucoma medications correlated with success.38 In our study previous ALT did not influence the success of SLT at 1 year.