Ultrasound velocity in heavy ocular tamponade agents and implications for biometry
  1. M A Rehman Siddiqui1,
  2. M Amer Awan1,
  3. Andrew Fairhead3,
  4. Hatem Atta2
  1. 1Tennent Institute of Ophthalmology, Gartnavel General Hospital, Glasgow, UK
  2. 2Department of Ophthalmology, Grampian University Hospitals NHS Trust, Aberdeen, UK
  3. 3The Department of Bio-Medical Physics and Bio-engineering, University of Aberdeen, UK
  1. Correspondence to Dr MA Rehman Siddiqui, Department of Ophthalmology, Grampian University Hospitals NHS Trust, Aberdeen AB25 2ZN, UK; rehman.siddiqui{at}gmail.com

Abstract

Aim Heavy ocular tamponade agents have been shown to be an effective tamponade in complicated retinal detachments. Combined oil removal and cataract surgery may be performed and ultrasound (US) provides a reliable means to measure the axial length of eyes. The aim of this study was to evaluate the velocity of US in various tamponade agents.

Methods Five tamponade agents (SO 1000, SO 5000, Oxane HD, Densiron 68 and F-Decalin) were studied in vitro. Time of flight (T1) was measured between a US transducer and the bottom of a container of the agent, and remeasured (T2) after reducing the separation by 1 cm. The speed of sound in the particular material was calculated from the difference between T2 and T1. Measurements were repeated over a range of temperatures from 18°C to 42°C.

Results The speed of sound at 37°C ranged from 645±8 m/s to 976±10 m/s depending on the tamponade agent. In Densiron 68 and F-Decalin, the speed of sound was markedly reduced to 914±10 m/s and 645±8 m/s, respectively. The temperature dependence of speed of sound varied between −2.2 and −3.6 m/s per °C depending on the particular oil. With 95% confidence, the true speed is believed to lie within ±5m/s of these values. The deviation from intended refraction was between 0.38 to 30.15 D depending on the tamponade agent by using correction factor (CF) for SO 1000 cs.

Conclusions Variability in the velocity of US should be taken into account when performing biometry. Biometry machines should be adjusted for various tamponade media when calculating intraocular lens power.

  • Biometry
  • heavy silicone oil
  • cataract surgery
  • lens and zonules
  • retina
  • optics and refraction

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Footnotes

  • Competing interests None to declare.

  • Provenance and peer review Not commissioned; externally peer reviewed.

Linked Articles

  • At a glance
    Harminder S Dua Arun D Singh