Article Text
Research Article
Tonometer disinfection and viruses.
Abstract
Tonometer disinfectants in current use are frequently inactive against viruses and enable transfer of infection between patients. We selected sodium hypochlorite as a readily available alternative which is active against viruses and bacteria. At 500 p.p.m. it destroys herpes simplex, adenovirus 8, and enterovirus 70 within 10 minutes. A simple technique is described which is clinically acceptable and has been in daily use for one year.